Combination Products Under Article 117 MDR: Regulatory Guidance for Manufacturers

Introduction

If your medicinal product includes a medical device as an integral component, compliance with Article 117 of the EU Medical Device Regulation (MDR 2017/745) is critical. Article 117 ensures that the device element of a combination product meets the same safety and performance requirements as a standalone medical device under the MDR.

For manufacturers preparing a Marketing Authorisation Application (MAA) under Directive 2001/83/EC or Regulation (EC) No 726/2004, Article 117 defines how conformity of the device component must be demonstrated as part of the medicines approval process. Failure to address these requirements appropriately can lead to delays, questions from competent authorities, or requests for additional assessment by a Notified Body.

Understanding Combination Products Under Articles 1(8) and 1(9) of the MDR

Before considering Article 117 in detail, it is important to understand how the MDR defines combination products under Articles 1(8) and 1(9).

We also recommend reading through MDCG 2022-5 on, “Guidance on borderline between medical devices and medicinal products under Regulation (EU) 2017/745 on medical devices”.

Article 1(8): Devices Containing a Medicinal Substance

Where a medical device incorporates a medicinal substance that has an ancillary action to that of the device, the product is regulated as a medical device under the MDR.

If, however, the medicinal substance has a principal action, the product is regulated as a medicinal product under Directive 2001/83/EC or Regulation (EC) No 726/2004. In these cases, the device component must still comply with the General Safety and Performance Requirements (GSPR) set out in Annex I of the MDR.

Article 1(9): Devices for Administering Medicinal Products

Where a device is intended to administer a medicinal product and is placed on the market separately, such as a reusable insulin pump, it is regulated as a medical device under the MDR.

Where the device and medicinal product form a single integral, non-reusable product, such as a prefilled syringe or auto-injector, the product is regulated as a medicinal product. However, the device component must still comply with the applicable GSPR requirements of Annex I of the MDR.

These scenarios represent the most common types of combination products affected by Article 117.

What Is Article 117 of the MDR?

Article 117 introduces a formal requirement for the device component of certain medicinal products to be assessed against MDR requirements as part of the MAA.

For medicinal products governed by Directive 2001/83/EC, where Articles 1(8) or 1(9) apply, the regulatory dossier must include evidence of conformity of the device component as follows:

  • Class I devices: An EU Declaration of Conformity issued by the manufacturer.
  • Higher-risk devices (Class Is, Im, Ir, IIa, IIb, III): A valid CE certificate issued by a Notified Body.

This requirement applies even though the overall product is authorised as a medicinal product rather than CE marked as a medical device.

When Is a Notified Body Opinion (NBOp) Required?

Where the device component is classified as higher risk and a CE certificate is not available, the competent authority will require a Notified Body Opinion (NBOp) as part of the MAA.

The NBOp confirms that the device component complies with the relevant GSPR in Annex I of the MDR. It effectively bridges the regulatory gap between medicinal product legislation and the MDR.

For Class I devices, where an EU Declaration of Conformity is not available, a statement from the Marketing Authorisation Holder confirming compliance with the relevant GSPR may be acceptable, subject to competent authority agreement.

Preparing a Notified Body Opinion Submission

A clear and well-structured submission is essential to avoid delays and additional questions during review.

Cover Letter

The cover letter should clearly outline the scope of the submission and typically include:

  • The NBOp reference number, if available
  • The type of submission, such as a new product, variation or line extension
  • A concise product and device description, including model numbers
  • Relevant Notified Body reference numbers for related activities
  • A justification of the scope of assessment and applicable MDR requirements

For variations or changes, it is important to clearly identify which elements of the device are affected and which remain unchanged, supported by an appropriate rationale.

Technical Documentation

The technical documentation should be structured in line with MDR Annex II and demonstrate conformity with the GSPR.

Device Description (Annex II, Section 1)
This section should include a description of the device, its variants and accessories, confirmation of the MDR classification, and a clear justification that the device falls within the scope of Article 117. It should also document that the product forms a single integral combination with the medicinal substance, is intended exclusively for that use, and is non-reusable.

Information Supplied by the Manufacturer (Annex II, Section 2)
Labelling and instructions for use must comply with medicinal product legislation. While GSPR 23 does not apply in full, information relating to device use, dosing accuracy, safety and risk communication must be addressed appropriately.

Design and Manufacturing (Annex II, Section 3)
This section should describe the design and manufacturing processes for the device, including all relevant manufacturing sites.

General Safety and Performance Requirements (Annex II, Section 4)
A GSPR checklist should be provided, demonstrating how each applicable requirement of Annex I is met, with clear references to supporting evidence.

Risk Management (Annex II, Section 5)
Risk management documentation should be provided in accordance with ISO 14971, including the Risk Management Plan, Risk Analysis and Risk Management Report.

Product Verification and Validation (Annex II, Sections 6.1 and 6.2)
Relevant pre-clinical testing should be included, such as mechanical, functional and biocompatibility testing.

Clinical Evaluation Considerations
In many Article 117 cases, the device is limited to drug delivery and does not require a standalone clinical evaluation. However, a clinical evaluation may be required where the device makes independent clinical claims, incorporates novel technology, or introduces specific safety risks, such as implantable components.

How We Support Combination Product Manufacturers

Our team provides end-to-end regulatory affairs and quality assurance support for combination products under Article 117. This includes device classification, MDR gap assessments, preparation of technical documentation, and strategic support for Notified Body Opinions.

Key Takeaways

  • Article 117 ensures that device components of combination products meet MDR safety and performance requirements.
  • Articles 1(8) and 1(9) determine whether a product is regulated under the MDR or medicinal product legislation.
  • A Notified Body Opinion is required where higher-risk device components lack a CE certificate.
  • Early regulatory planning, robust documentation and experienced regulatory support are essential to avoid delays in the MAA process.