BCWG Publishes Borderline Manual Version 5: Five New Entries Manufacturers Should Review

The Borderline and Classification Working Group (BCWG), a subgroup of the Medical Device Coordination Group, has published Version 5 of the Manual on Borderline and Classification for Medical Devices under Regulation (EU) 2017/745 (MDR) and Regulation (EU) 2017/746 (IVDR). Released in April 2026, this update adds five new case entries to the manual, each reaching a conclusion through the Helsinki Procedure.

For manufacturers, this manual is one of the most practical regulatory reference tools available. It records the agreed positions of Member State competent authorities on borderline cases; products where it is not immediately clear whether they qualify as medical devices, IVDs, or something else entirely, and where questions about the appropriate classification rule have arisen. The decisions are not legally binding in themselves, but they represent the coordinated view of national authorities across the EU.

Here is what is new in Version 5.

Saline solutions for nasal irrigation: Class IIa under Rule 21

This entry will likely attract the most attention. Saline nasal irrigation solutions, the 0.9% sodium chloride products used to relieve nasal congestion and treat rhinosinusitis, are confirmed as Class IIa medical devices under Rule 21 of Annex VIII to the MDR.

Some manufacturers had been placing these products on the market as Class I devices under Rule 5, which applies to invasive devices introduced into body orifices. The BCWG rejected this approach. The saline solution is a substance-based device that achieves its intended purpose within the nasal cavity, and the guidance MDCG 2021-24 specifically cites salt water nose sprays as examples of Class IIa products. Rule 21, which covers devices composed of substances applied in the nasal or oral cavities, is the applicable rule.

This matters for any manufacturer currently selling these products as Class I or outside MDR scope entirely. Class IIa requires conformity assessment involvement from a Notified Body. If your technical documentation and QMS were built around a Class I assumption, a classification review is now overdue.

Root canal irrigation needles: Class IIa under Rule 6

Single-use flexible needles used during endodontic (root canal) procedures are confirmed as Class IIa under Rule 6, which applies to short-term surgically invasive devices.

The classification question here turned on whether access to the root canal constitutes a surgical intervention. Several manufacturers had classified these needles under Rule 5, applying to non-surgical invasive devices introduced through body orifices, resulting in a Class I designation. The BCWG disagreed. Reaching the root canal requires drilling an opening in the tooth, an artificially created access point, not a natural or permanent body orifice. Since guidance MDCG 2021-24 specifies that surgical invasiveness applies when a device enters the body through an artificially created opening, Rule 6 governs these devices. With contact duration under 60 minutes, the classification is Class IIa.

Any manufacturer who has been selling these needles as Class I will need to revisit their regulatory strategy.

Syringes containing glass beads for ACS preparation: Class IIb under Rule 3

The BCWG addressed the classification of syringes containing borosilicate glass beads, designed for the collection, incubation and centrifugation of blood to produce autologous conditioned serum (ACS), a concentration of growth factors used in treatments such as knee osteoarthritis.

The glass beads increase surface area to accelerate coagulation via contact stimulation. The device is non-invasive but modifies the composition of blood intended for administration back into the patient. Rule 3 of Annex VIII to the MDR governs devices that modify biological constituents intended for administration into the body. The first paragraph of Rule 3 applies, classifying the syringe system as Class IIb. The Class IIa exception for simple filtration, centrifugation or gas/heat exchange does not apply here, as the mechanism is more complex.

Rectal administration packaging as a medical device: Article 117 MDR

A laxative product packaged as a single-dose mini tube with a long nozzle for rectal insertion was considered. The question was whether this delivery packaging qualifies as a medical device.

A device invasive to a body orifice for the purpose of administering a medicinal product falls within the MDR’s definition of a medical device. However, when the device and medicinal product are placed on the market as a single integral combination, not intended for reuse, the product is governed by Directive 2001/83/EC or Regulation (EC) No 726/2004, as applicable. In this situation, Article 117 MDR requires the device component to comply with the relevant general safety and performance requirements of Annex I to the MDR. Proof of conformity must be included in the marketing authorisation dossier for the combined product.

This entry is relevant to pharmaceutical companies whose medicinal products incorporate device packaging for administration into body orifices.

Penis holster for urinary incontinence: Class I under Rule 1

A device for men with urinary incontinence that transports urine from the urinary meatus to a collection bag was classified as Class I under Rule 1.

The question concerned whether the device, which sits under the inner mucosa of the prepuce, constitutes invasive contact through a body orifice. Under the MDR and MDCG 2021-24, a body orifice means any natural opening in the body, the external surface of the eyeball, or any permanent artificial opening such as a stoma. The inner mucosa of the prepuce does not fall within this definition. The device is therefore not invasive through a body orifice and is appropriately classified as Class I under Rule 1.

What this means for your product portfolio

Each of these entries reflects a genuine disagreement between manufacturers and the competent authorities that had to be resolved through a coordinated EU-level process. They are not edge cases. Nasal saline sprays and root canal needles are common, widely sold products. The fact that classification disputes exist for these products is a reminder that regulatory assumptions, particularly for devices with a long history on the market, deserve periodic review.

If any of the product types covered in this update appear in your portfolio, the next step is to compare your existing classification rationale against the BCWG’s reasoning and assess whether your Notified Body, technical documentation, and QMS reflect the correct classification.

The full manual is available on the European Commission’s MDCG website.

If you would like to discuss how any of these entries affect your device classification, we are happy to help. Book a call with our team to talk through your specific situation: info@specculo.com