On 20 March 2026, the European Commission adopted two delegated regulations expanding the Well-Established Technology (WET) lists under the MDR. This article covers C(2026) 1809, which expands the list of Class IIb implantable devices exempt from the requirement for a notified body to assess the technical documentation for every individual device.
The companion regulation, C(2026) 1798, addresses a separate WET exemption for implantable and Class III devices in the context of clinical investigations and is covered in a separate article.
Both regulations were adopted on 20 March 2026 and are not yet published in the Official Journal of the European Union. Both enter into force 20 days after OJEU publication.
The Conformity Assessment Burden for Class IIb Implantable Devices
Under Article 52(4) MDR, Class IIb implantable devices are subject to conformity assessment via Annex IX, which includes an assessment of technical documentation by the notified body. The default rule is that this technical documentation assessment applies to every device in the product range.
For manufacturers with a broad portfolio of Class IIb implantable devices, this creates a significant workload. A notified body assessing each device individually across a large product family generates both time and cost implications at initial certification and at renewal.
Article 52(4) already contained a limited exemption for specific device types listed in the regulation itself. Article 52(5) empowers the Commission to expand that list by delegated act. C(2026) 1809 is the first exercise of that power to add a meaningful number of new device types.
What C(2026) 1809 Changes
The regulation replaces the second subparagraph of Article 52(4) with an updated list. For the device types on this list, the notified body is not required to assess the technical documentation for every individual device. A single representative assessment of the device type covers the product family.
The updated list is structured in two parts.
Category (a) retains the original devices: sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors.
Category (b) adds a new set of device types: cannulas, catheters, feeding tubes, suture pledgets, suture sleeves, suture buttons, gastrostomy buttons, bone wax, bone fillers, bone substitutes, stem centralisers, diaphyseal obturators, radiography markers, fiber ligatures, transpalatal distractors, nails, anchors, spinal posterior fixations, textile braids, dental implants, orthodontic devices, dental barriers, suspensory fixations and cinches.
Note that this list is specific to Class IIb implantable devices. The parallel regulation C(2026) 1798 covers clinical investigation exemptions across both implantable and Class III devices and contains a longer list of device types including several that do not appear here.
What the Exemption Actually Means
Where a Class IIb implantable device type is on the Article 52(4) list, the conformity assessment process changes in one specific respect: the notified body conducts a representative technical documentation assessment rather than an individual assessment for every device variant.
In practical terms, this means:
Reduced NB scope per product family. A manufacturer with a range of, say, dental implants in varying dimensions and materials does not need each individual configuration assessed separately. The NB reviews representative devices and extrapolates to the broader family, provided the variants are genuinely similar in design and intended purpose.
Lower conformity assessment cost and shorter timelines. Fewer individual assessments mean less NB time, which typically translates into lower fees and a more streamlined certification cycle.
The technical documentation itself is still required. The exemption covers the NB’s obligation to assess that documentation for every device but it does not remove the manufacturer’s obligation to compile complete and compliant technical documentation for each device. The documentation must still exist and be available to the NB; the question is only whether the NB must actively review it for each variant.
Scope: Class IIb Implantable Devices Only
It is worth being clear on the scope. This regulation applies exclusively to Class IIb implantable devices. It does not apply to:
- Class IIb devices that are not implantable
- Class III devices (those are covered by a separate framework under Article 61)
- Class IIa or Class I devices
If your device is Class IIb but is not implantable, this regulation does not apply. The classification question is therefore the starting point for any assessment of eligibility.
Comparing the Two WET Regulations
It is also worth noting that the two lists in C(2026) 1798 and C(2026) 1809 are related but not identical, and they address different compliance obligations. A device that appears on both lists benefits from both exemptions: no mandatory clinical investigation under Article 61(6)(b), and no per-device TDoc assessment by the NB under Article 52(4). A device that appears on one list but not the other benefits only from the corresponding exemption.
Manufacturers should check both lists carefully and assess the combined impact on their compliance strategy.
Timing and Next Steps
Both regulations were adopted on 20 March 2026. Neither has yet been published in the OJEU. The 20-day entry-into-force clock starts on OJEU publication. Until that date, the current Article 52(4) list remains in effect.
For manufacturers of Class IIb implantable devices, the practical steps are:
First, check the updated list in C(2026) 1809 and confirm whether your device type is included. The list describes device types at a category level, so you will need to assess whether your specific products fall within a qualifying category.
Second, if your device qualifies, review your current NB agreement and conformity assessment scope. The exemption does not automatically amend an existing NB contract. You will need to raise this with your notified body and agree on a revised assessment approach.
Third, consider the combined picture. If your device also appears on the C(2026) 1798 list under Article 61(6)(b), you may be in a position to revise both your clinical investigation plans and your NB assessment scope. That is a meaningful change to your compliance roadmap and worth working through carefully.
If you are not sure whether your device qualifies, or what the practical impact is on your certification timeline and NB relationship, we are happy to help you think it through. Get in touch.