EMA Breakthrough Devices Pilot: Phase Ib Is Now Open to Class III and Certain Class IIb Devices

On 1 September 2026 the European Medicines Agency opened Phase Ib of its expert panel support for breakthrough medical devices pilot programme. Manufacturers of Class III devices, and of Class IIb active devices intended to administer or remove a medicinal product, can now submit a letter of interest asking the expert panels for an opinion on breakthrough status.

In summary, phase Ia was a small, competitive first round with a fixed deadline. Phase Ib is the open round: EMA says it “accepts all applications within the scope of phase I”, there is no deadline for the letter of interest, and EMA is setting up weekly submission slots until the end of 2026. The advice is free.

We covered the launch of the pilot in May in EMA Breakthrough Devices Pilot Programme: What Manufacturers Need to Know. This post picks up where the pilot stands now.

Where the pilot stands

EMA launched the pilot on 28 April 2026 and says it will run in three phases until 2027, subject to available resources. Its stated aim is to establish a long-term process for manufacturers and notified bodies to request support from the medical device expert panels.

Phase Ia is closed for new submissions. Manufacturers selected in Phase Ia who received a positive opinion on the breakthrough status of their device can submit requests for advice from 1 September 2026. EMA also notes that applicants not selected for Phase Ia but considered eligible for Phase Ib have already received a communication on next steps.

The scope for Phase I as a whole, and so for Phase Ib, is limited to:

  • Class III medical devices; and
  • Class IIb active devices intended to administer or remove a medicinal product.

How to apply

Letters of interest go through EMA’s dedicated portal, the Business Services Employee Center. When submitting, select the form titled “Breakthrough device according to MDCG guidance 2025-9”. EMA is working through submissions in weekly slots up to the end of 2026.

The timeline from here

EMA has published the following expected phases:

  • Phase Ib: open from 1 September 2026, weekly submission slots until the end of 2026.
  • Phase II: expected to start in Q1 2027. EMA extends designation and advice requests to all classes of medical devices.
  • Phase III: expected to start in Q3 2027. EMA extends designation and advice requests to in vitro diagnostic medical devices.

The Phase II and Phase III dates are expectations, and EMA frames the whole pilot as subject to available resources, so best to treat them as planning assumptions.

What counts as a breakthrough device

The criteria come from MDCG 2025-9, published in December 2025 and covered in our earlier post on the MDCG breakthrough devices guidance. A device must meet both of two conditions.

First, novelty. The device must introduce a high degree of novelty with respect to the device technology, the related clinical procedure, and/or the application of the device in clinical practice.

Second, positive clinical impact for a life-threatening or irreversibly debilitating disease or condition. That impact can be shown either as a significant positive clinical impact on patients or public health compared with available alternatives and the state of the art, or by fulfilling an unmet medical need where alternative options are absent or insufficient.

Both parts matter and a device can be technically novel and still fall short if the condition it addresses is not life-threatening or irreversibly debilitating, or if the benefit over the state of the art is incremental.

What support is on offer

Through the pilot, the expert panels can provide an opinion on the breakthrough status of a device. For designated breakthrough devices, they can then advise on specific areas, which EMA lists as pre-clinical evidence generation, the acceptability of the intended clinical development strategy, clinical investigations, and post-market clinical follow-up activities. Notified bodies can also submit specific questions to the expert panels on implementing the guidance.

In practice, this gives you an expert view on your clinical evidence plan before you commit to a clinical investigation, while changes are still relatively inexpensive to make.

What this means for manufacturers

If you have a Class III device, or a Class IIb active device that administers or removes a medicinal product, and you think it could meet the MDCG 2025-9 criteria, Phase Ib is the lowest-friction entry point the pilot has offered so far. There is no competitive cut-off and no fixed deadline, although the weekly slots are only scheduled to the end of 2026.

Before submitting, we would suggest three steps:

  1. Test the device honestly against both criteria. Write down the novelty claim and the clinical impact claim separately, with the state of the art you are comparing against. If the clinical impact case rests on a condition that is serious but not life-threatening or irreversibly debilitating, that is worth resolving first.
  2. Prepare the evidence position. Advice is most useful when you can show what pre-clinical and clinical data you already hold and what you plan to generate. A clear draft clinical development plan gives the panel something concrete to respond to.
  3. Line up your notified body. The pilot is designed to involve notified bodies as well as manufacturers. Letting your notified body know you are applying keeps later discussions aligned.

If your device sits in a lower class or is an IVD, the Phase II (Q1 2027) and Phase III (Q3 2027) windows are the ones to plan for. The same preparation applies, and starting it now means you are ready when those phases open.

Specculo supports manufacturers of high-risk and innovative devices with regulatory strategy and clinical evidence planning, including preparing for expert panel interactions. If you are weighing whether your device has a credible breakthrough case, get in touch and we can talk it through.