The European Medicines Agency launched a pilot programme on 28 April 2026 offering manufacturers of designated breakthrough medical devices and IVDs access to priority scientific advice from EMA expert panels. An information session was held on 24 April 2026, ahead of the pilot’s opening. The Phase Ia application deadline is 22 May 2026.
This is worth understanding properly. The programme is not just an administrative novelty. For manufacturers developing genuinely novel devices in indications with high unmet need, earlier expert panel engagement could meaningfully reduce the risk of late-stage clinical evidence challenges.
The two foundations
The pilot builds on two things that are worth distinguishing clearly.
The first is MDCG 2025-9, published in December 2025, which established the framework for breakthrough device designation under the existing MDR and IVDR. MDCG 2025-9 sets out the criteria a device must meet to be considered breakthrough, the designation process, and what designation actually entitles a manufacturer to. The pilot programme operationalises the expert panel engagement component of that framework.
The second is the proposed reform package from December 2025 (COM/2025/1023), which includes proposed new Articles 52a (MDR) and 48a (IVDR). These proposed articles would give the breakthrough pathway a formal legislative basis within the regulations themselves. They are still moving through Parliament and Council. The pilot operates on the MDCG 2025-9 framework in the meantime, so you don’t need to wait for the reform to progress before engaging with the programme.
What breakthrough designation actually means
A device is eligible for breakthrough designation under MDCG 2025-9 if it is intended to treat or diagnose a serious or life-threatening condition; if there is no suitable alternative therapy or diagnostic currently available to patients, or the device offers a clinically meaningful advantage over the standard of care in terms of safety, clinical performance, or patient outcomes; and if there is sufficient preliminary clinical evidence to support a plausible claim of advantage.
Two things are worth stating clearly here. Designation is not a classification decision. It does not change which conformity assessment procedure applies or what clinical evidence standard the notified body will apply. A Class III implantable with breakthrough designation is still a Class III implantable, assessed under Annex IX or XI MDR, with full compliance with Annex XIV clinical evaluation requirements. What designation changes is the quality and timing of engagement available before you finalise your evidence strategy.
What the pilot offers in practice
Manufacturers with breakthrough designation gain access to priority scientific advice from EMA expert panels at earlier stages of clinical development than the standard scientific advice route typically allows.
The practical benefit sits in a specific and well-understood problem: a significant proportion of device clinical trial failures, or late-stage notified body challenges, arise from questions that were answerable before the trial started. Whether the chosen primary endpoint is considered clinically meaningful. Whether a surrogate endpoint is acceptable given the indication and patient population. Whether a specific comparator was appropriate. Whether the evidence generation plan as a whole is proportionate and persuasive.
These are not questions that become easier to answer once the trial is running. Earlier expert panel engagement allows them to be resolved at the point where they are cheapest to address: before the protocol is finalised, before sites are contracted, and before patients are enrolled.
The pilot structure
The pilot runs in phases through to 2027, subject to available resources.
Phase Ia accepts five applications. The application deadline is 22 May 2026. Phase Ib follows in Q3 2026 and opens to manufacturers of high-risk medical devices, specifically Class III and Class IIb active devices intended to administer or remove a medicinal product. All applications within scope are accepted in Phase Ib.
The scope in Phase Ia is not limited to those device types, but the selection is competitive given the cap of five places. Manufacturers with strong preliminary evidence and a well-articulated unmet need argument are best positioned.
Who should be paying attention
The programme is most immediately relevant for three groups.
Manufacturers developing Class III implantables and active implantable devices in indications with limited existing evidence bases, where the clinical evidence strategy is genuinely novel territory. Class IIb implantables with novel mechanisms of action in high-unmet-need indications. And IVD manufacturers in Class C and D where clinical utility evidence is still maturing at the point of development.
It is also worth considering if your clinical evidence strategy is likely to face scrutiny at technical documentation review stage, regardless of device class. Earlier EMA panel alignment on endpoints and study design reduces the probability of late-stage notified body challenges, which can add 18 to 24 months to a certification timeline. That is a concrete risk reduction, not a regulatory shortcut.
What it does not change
The pilot does not accelerate the conformity assessment timeline. It does not reduce clinical evidence requirements. Notified bodies will still apply the General Safety and Performance Requirements in Annex I of MDR and IVDR, and the clinical evaluation requirements in Annex XIV (MDR) and Annex XIII (IVDR), in full. A breakthrough designation is not a licence to submit a lighter technical file.
What it changes is the probability that your clinical evidence, gathered over time and at significant cost, actually satisfies the requirements on first review.
As the reform progresses
The proposed Articles 52a and 48a are worth monitoring as they move through the legislative process. If they are adopted, the breakthrough pathway will move from MDCG guidance into primary law, with greater certainty about scope, criteria, and process. The pilot is the practical early version of what could become a standing programme.
For now, if you are developing a device that plausibly meets the MDCG 2025-9 criteria and the Phase Ia deadline is live, the question is worth asking seriously.
Thinking about whether your device qualifies?
The breakthrough designation criteria are more specific than they first appear, and the case for eligibility needs to be made clearly to be competitive in Phase Ia. If you’d like to think through whether your device meets the MDCG 2025-9 threshold and how to frame the application, we’re happy to work through it with you. Get in touch.
References