Two EU digital obligations took effect within a day of each other this month. From 12 September 2026, the access by design requirement in Article 3(1) of the Data Act, Regulation (EU) 2023/2854, applies to connected products and related services placed on the market after that date. The day before, on 11 September 2026, manufacturers’ vulnerability and incident reporting duties under the Cyber Resilience Act began to apply, and ENISA launched the Single Reporting Platform that carries them.
These two instruments pull in opposite directions for medical device manufacturers. The Data Act has no general exclusion for devices, so a connected device placed on the market from now on is likely to be in scope. The CRA does exclude products covered by the MDR or IVDR, but that exclusion follows the product rather than the portfolio, so the apps, hubs and accessories sold around a regulated device may sit outside it. Both questions have to be answered product by product.
What the Data Act now requires
Article 3(1) requires connected products and their related services to be designed so that the data they generate, together with the metadata needed to interpret and use it, is available to the user by default. That data must be easy to get at, secure, free of charge, and in a comprehensive, structured, commonly used and machine readable format. Where it is relevant and technically feasible, the user should be able to reach it directly from the product.
Article 50 sets the timing. The Regulation has applied since 12 September 2025, but the Article 3(1) design duty attaches only to connected products and related services placed on the market after 12 September 2026. Because it is a design duty, it has to be settled during development. Retrofitting access into a product that is already finished is the expensive way to comply.
The rest of Chapter II already applies, and applies to products that were on the market before that date. The user’s right to ask for data that is not directly accessible, under Article 4, and to have it shared with a third party of their choosing, under Article 5, have both been live since 12 September 2025. So the September 2026 date does not start the obligation. It adds a design duty on top of access rights that are already running.
Whether your device is caught
On the text of the Regulation, most connected devices are. The Data Act does not carve out products regulated under the MDR or IVDR, and Recital 14 lists “medical and health devices” among its examples of connected products, alongside “health and lifestyle equipment”. A device that collects data about its own performance, use or environment and can communicate that data is likely to be a connected product whatever its risk class.
Four points shape how far the duty actually reaches.
What counts as data. The obligation covers product data and related service data together with the metadata needed to interpret it. Recital 15 indicates that information derived or inferred from that data through further processing falls outside the definition. Where that line sits for a particular device, say between a raw sensor reading and the output of an algorithm applied to it, is a judgement each manufacturer has to make and record.
Relevance and feasibility. The qualifier “where relevant and technically feasible” governs direct access only. Where direct access is not feasible, the Article 4 route of supplying data on request still stands, so the qualifier changes the mechanism rather than removing the duty.
Safety. Article 4(2) allows users and data holders to agree contractually to restrict or prohibit access, use or onward sharing where that processing could undermine security requirements of the connected product laid down in Union or national law, resulting in a serious adverse effect on the health, safety or security of natural persons. A data holder that refuses to share data on this basis has to notify the competent authority. The MDR and IVDR security requirements are the obvious reference point for a device, though how far manufacturers can lean on this has not been tested.
Company size. Under Article 7(1), the Chapter II obligations do not apply to data generated by connected products manufactured or designed by a microenterprise or small enterprise, subject to conditions covering partner and linked enterprises and subcontracting. Time limited relief also covers companies that have recently grown into the medium sized bracket. Plenty of device software companies will qualify, and it is worth checking before assuming the duty applies.
Much device data will be health data under the GDPR, which continues to apply in full. And a change to software or interfaces made in order to provide data access is still a design change, so it needs to go through MDR or IVDR change control and, where the change is significant, past your notified body.
Where the EHDS fits
The European Health Data Space Regulation, Regulation (EU) 2025/327, governs electronic health data, EHR systems and wellness applications claiming interoperability with them, and is being phased in over the coming years. It and the Data Act cover different subject matter, and we would not assume either displaces the other. A connected device whose data feeds an EHR system, or a companion app claiming EHR interoperability, may well need assessing under both. How the two interact in practice is unsettled, and further Commission guidance would help.
The CRA reporting duties that started on 11 September
Most of the Cyber Resilience Act, Regulation (EU) 2024/2847, applies from 11 December 2027 under Article 71. Article 14 is the exception and has applied since 11 September 2026. Article 69(3) extends those reporting duties to all in scope products with digital elements, including products placed on the market before December 2027, so there is no grace period for an existing portfolio.
Article 14 requires manufacturers to notify actively exploited vulnerabilities and severe incidents affecting the security of their products, through the Single Reporting Platform, to the coordinating CSIRT and to ENISA. The clocks are short:
- Actively exploited vulnerabilities: early warning within 24 hours of becoming aware, a fuller notification within 72 hours, and a final report within 14 days of a corrective or mitigating measure becoming available.
- Severe incidents: early warning within 24 hours, notification within 72 hours, and a final report within one month of that notification.
Twenty-four hours is a duty operations teams have to be staffed and rehearsed for. A quarterly review process will not catch it.
Why devices are excluded, and what is not
Article 2(2) provides that the CRA does not apply to products with digital elements covered by certain listed Union acts, and both Regulation (EU) 2017/745 and Regulation (EU) 2017/746 are on that list. A device regulated under the MDR or IVDR, medical device software included, therefore sits outside the CRA, and its cybersecurity continues to be governed by those Regulations and their vigilance system.
The exclusion attaches to the individual product, and connected device ecosystems usually contain products that are not themselves regulated devices. Depending on how each one qualifies, the following may fall inside the CRA:
- a general wellness app, or any app without a medical intended purpose, that works alongside the device;
- companion hardware such as a hub, gateway or charging dock that is neither a device nor an accessory within the meaning of the MDR or IVDR;
- consumer facing software or cloud connected products supplied with, but separate from, the regulated device.
Qualification decides each case and the borderline ones need reasoning through properly. Where a product does fall inside the CRA, the reporting duties apply now, even though the essential cybersecurity requirements do not arrive until December 2027.
What to do next
We would treat this as one scoping exercise across the whole connected portfolio rather than two separate compliance projects.
List every product in the ecosystem and record its status. The device, its software, companion apps, hubs, cloud services and accessories, each marked as MDR, IVDR or neither. Everything below depends on this list being accurate.
Check Data Act scope for each connected product and related service. Identify the product data and related service data it generates, note which products will be placed on the market after 12 September 2026, and establish whether the Article 7 exemption covers your company before doing further work.
Decide and document the access design. For products still in development, record how data will reach users by default, where direct access is not relevant or technically feasible and why, and the justification for any Article 4(2) restriction you intend to rely on.
Route access changes through device change control. Treat any design change made for data access as a change under your MDR or IVDR procedures, and involve your notified body where the change is significant.
Check CRA scope for everything outside the MDR and IVDR. Where a product is in scope, name an owner for reporting, get them access to the Single Reporting Platform before you need it, and confirm the 24 hour clock can actually be met out of hours.
Recording your scoping decisions and the reasoning behind them now is what will make them defensible when an authority or a customer asks.
Specculo helps connected device and software manufacturers work through qualification, classification and regulatory strategy across the MDR, IVDR and the wider EU digital framework. If a product by product scoping check would be useful, get in touch or read more about our regulatory consulting services.