Manufacturers are required to summarise the results and conclusions of the post-market activities which have been conducted in accordance with a PMS Plan and record a rationale and description of any further corrective or preventive actions which may need to be taken. This shall be done for each device, category or group of devices for a given reporting interval.
For Class I medical devices and Class A and B IVDs, the report takes the form of a Post-Market Surveillance Report. Manufacturers of higher risk devices, including Class IIa, IIb, III medical devices and Class C, D IVDs are required to put together a Periodic Safety Update Report (PSUR).

Summary
PMS Report & PSUR Content Overview
| PMS Report | PSUR | |
| Classifications | MDR: Class I IVDR: Class A, B | MDR: Class IIa, IIb, III IVDR: Class C, D |
| Content | ||
| Summary of Results of PMS Activities | X | X |
| Conclusions of Analyses of PMS Activities | X | X |
| Rationale & Description of Corrective or Preventive Actions to be taken | X | X |
| Conclusions of Benefit-Risk Determination | X | |
| Main findings of PMCF | X | |
| Sales Volume | X | |
| Estimate of Size & Other Population Characteristics | X | |
| Estimate of Usage Frequency, where applicable | X |
PMS Report & PSUR Update Frequency
| When Necessary | Every 2 years (at least) | Annually (at least) | |
| MDR: Class I IVDR: Class A, B | X | ||
| MDR: Class IIa | X | X | |
| MDR: Class IIb, III IVDR: Class C, D | X | X |
Report Availability
| Competent Authority (when requested) | Notified Body | Eudamed | |
| MDR: Class I IVDR: Class A | X | ||
| MDR: Class IIa IVDR: Class B, C | X | X | |
| MDR: Class III MDR: Implantable IVDR: Class D | X | X | X |
Post-Market Surveillance Report
The PMS Report is described briefly, with very few specifics, in Article 85 and Article 80 of the EU MDR and IVDR, respectively. Data resulting from post-market activities shall be collected and assessed within the report. The PMS Report shall be provided to the Competent Authority, upon request and should therefore be updated at regular intervals.
Periodic Safety Update Report
On the other hand, the expected content of the PSUR is described in Article 85 and Article 81 of the EU MDR and IVDR, respectively. The overall concept remains the same: summarise the results of post-market activities and justify any further actions being taken.
The manufacturer is expected to provide a PSUR for each device, and where relevant for each category or group of devices. Throughout the product’s lifetime, the PSUR shall set out:
| (a) the conclusions of the benefit-risk determination; | Has the data collected affected the overall benefit-risk? – New information gathered in the post-market phase as more patients or users use the device. – A rare adverse event may not have been identified during clinical trials. – New data from PMCF demonstrating benefits to new indications or measuring risks. |
| (b) the main findings of the PMCF; and | Discuss the main findings of the PMCF study conducted. |
| (c) the volume of sales of the device and an estimate evaluation of the size and other characteristics of the population using the device and, where practicable, the usage frequency of the device. | – Provide the sales data for the given investigation period. – Estimate the population using the device (size, demographic, other). This can be tricky for reusable devices and should be calculated taking into account how often the device may be used on average. – Estimate the usage frequency of the reusable device. |
The PSUR will need to be provided to the Notified Body as part of their conformity assessment procedure. Manufacturers of Class III and implantable medical devices, as well as Class D IVDs will be required to upload the latest PSUR to Eudamed, when this function becomes available. Following review and acceptance, the Notified Body will electronically confirm the PSUR within the Eudamed system.
How should the PMS Report or PSUR be updated?
The PMS Report and PSUR are intended to be single, stand-alone documents for a particular reporting interval, based on cumulative data. This means that for every reporting interval, the manufacturer shall take into account the data collected for that particular period, however, this shall be added to the pool of data already collected. The benefit-risk needs to be determined based on all acquired data, present and past.
Structure of the PMS Report and PSUR
With an absence of formal guidance on the structure of the reports, we recommend the following:
| Section | PMS Report | PSUR |
|---|---|---|
| Reporting Interval | X | X |
| Introduction & Scope | X | X |
| Objectives | X | X |
| Device Description | X | X |
| Implementation of PMS Plan | X | X |
| Summary of Activities Conducted, their Results, and Conclusions | X | X |
| PMCF Findings | X | |
| Sales Volume | X | |
| Use Evaluation | X | |
| Benefit-Risk Determination | X | |
| Corrective or Preventive Actions | X | X |
| Technical Documentation Amendments | X | X |
How can Specculo help?
The complexity of authoring the PSUR, especially the Benefit-Risk and PMCF conclusions, often consumes significant internal regulatory resources. Failure to meet the mandatory frequency or content requirements results in immediate Notified Body non-conformities.
Specculo specialises in translating raw PMS data into audit-ready PSUR and PMS Reports, ensuring complete compliance with MDR Articles 85 and 86. Our RA/QA consulting services provide:
- PSUR Authorship: Expert drafting of your PSUR, focusing on the rigorous Benefit-Risk Determination and accurate PMCF findings required by your Notified Body.
- PMS System Design: Implementation or gap assessment of your PMS Plan to ensure sufficient data collection (vigilance, literature, complaints) is available to feed the reports.
- Data Integration Strategy: Guidance on harmonising PMS data with your Risk Management and Clinical Evaluation files, ensuring traceability for auditors.
Contact us today to streamline your MDR/IVDR PSUR and PMS Report compliance.