Expert Panel Opinion on a Class III Knee Injectable: Claims Must Wait for Clinical Data

On 22 September 2026 the European Commission published an expert panel opinion (reference CECP-EMA/EX/0000335537) on a Class III resorbable device injected into the knee for mild to moderate osteoarthritis. The panel for orthopaedics, traumatology, rehabilitation and rheumatology gave its view under the clinical evaluation consultation procedure, and its recommendations extend to how the device is marketed.

The short version is this. The panel told the notified body that two performance claims, supported only by pre-clinical data, should not appear in the instructions for use, online or in any promotional material until clinical data supports them. It also set out detailed expectations for post-market clinical follow-up.

The procedure in brief

Under Article 54 and Annex IX, Section 5.1 of the MDR, a notified body assessing a Class III implantable device, or a Class IIb active device intended to administer or remove a medicinal product, must consult an expert panel before issuing a certificate, unless an exemption applies. The notified body sends its clinical evaluation assessment report to the Commission, and the panel decides whether to give a scientific opinion. If it does, the notified body must give due consideration to that opinion and justify in its final report any decision to depart from it.

In this case the procedure started on 10 March 2026, the panel decided on 16 March 2026 to provide an opinion, and the opinion itself is dated 8 May 2026.

What the panel found

The device is a liposomal suspension with a lipid-polymer stabiliser, given as a single intra-articular injection. The pivotal trial was a double-blind, randomised, placebo-controlled study with a 26-week follow-up and a primary endpoint based on a KOOS pain subscore.

The panel acknowledged the notified body’s assessment but stated that “the critical appraisal of the manufacturer’s claims was insufficient.” It found the clinical benefit data inconclusive, noting that the confidence intervals for pain reduction crossed the minimal clinically important difference and in some cases crossed zero. It also flagged foreign body synovitis as a risk linked to the novel composition that had been evaluated only in preclinical testing, and found the notified body’s benefit-risk determination “not comprehensive.”

The two claims

The two claims in question were that the device “reduces wear to enable smooth motion” and “provides a long-lasting mechanical ‘cushion’ to enable ease of movement”. The panel’s recommendation to the notified body reads:

“Ensure that the two promotional claims (reducing wear to enable smooth motion; providing a long-lasting mechanical “cushion” to enable ease of movement) are not made in the IFU, online or in any promotional material until demonstrated and supported by clinical data.”

This is worth noting because it reaches beyond the technical documentation. The panel treated the website and promotional material as part of what the notified body should control.

That is consistent with the MDR. Article 7 prohibits text, names, pictures and other signs in the labelling, instructions for use, making available, putting into service and advertising of a device that could mislead the user or patient about its intended purpose, safety and performance, including by ascribing functions and properties the device does not have. Claims are therefore part of what the clinical evaluation must support, and they are assessed as part of conformity assessment.

The PMCF expectations

The panel’s recommendations for post-market clinical follow-up are specific:

  • Imaging to test the claims. “Add a MRI (cartilage specific) at baseline and at minimum 1-year post-treatment to the PMCF study to evaluate these claims.”
  • Confounders. “Co-medications / co-treatments (e.g. physiotherapy, diet etc.) during follow-up in both groups are essential variables to record.”
  • A specific risk. “The risk of foreign body synovitis should be further characterised and specifically evaluated.”
  • Repeat use. The PMCF plan should “clearly specify whether it is planned to study repeated injections” and at what frequency, linked to the synovitis risk.
  • Population. “It should be ensured that the device’s intended use and population are not expanded before sufficient data is collected.”

The panel also recommended a longer randomised controlled trial, preferably with three arms comparing the device with the current state of the art and with placebo, for at least one year.

What this means for manufacturers

If your device falls under the consultation procedure, assume the panel will read your claims as closely as your clinical data. Before your notified body sends its assessment report to the Commission, we suggest the following:

  1. Build a claims matrix. List every performance and benefit claim across the IFU, label, website, brochures, sales training and social media, and map each one to the clinical evidence in your clinical evaluation. Where a claim rests only on bench or animal data, either remove it or ensure you can back it up with appropriate clinical data.
  2. Bring marketing into the review. Promotional copy is often written after the technical file is closed. Your regulatory team should sign off on claims wherever they appear.
  3. Design PMCF to answer open questions. If a claim or a residual risk is not yet supported clinically, your PMCF plan should say how you will generate that data, with endpoints that measure it directly.
  4. Record confounders from the start. Co-treatments such as physiotherapy are easy to capture at enrolment and very hard to reconstruct later.
  5. Hold your intended population steady. Expanding indications before the data exists is likely to draw attention, both from the panel and your notified body.

These points apply well beyond orthopaedic injectables. Any manufacturer whose device goes through the consultation procedure should expect the same scrutiny of claims.

Specculo helps manufacturers align clinical evaluations, PMCF plans and product claims before a notified body or expert panel reviews them. If a claims review ahead of your consultation would help, get in touch or read more about our regulatory consulting services.