On 29 September 2026 the European Commission published an expert panel opinion (CECP-EMA/EX/0000171363) on a Class III bioabsorbable bone screw made from a degradable magnesium alloy. The opinion was given under the clinical evaluation consultation procedure in Article 54 of the MDR, by the expert panel on orthopaedics, traumatology, rehabilitation and rheumatology.
The short version is that the panel did not accept that one single-centre study in one fracture type could support a device intended for fracture fixation and osteotomy across the upper and lower limb. It asked for a pre-authorisation clinical study across indication groups, a comparative post-authorisation study with set imaging timepoints, and a clinical investigation within the PMCF plan. This is the second orthopaedic opinion published in a week, following the opinion on a Class III knee injectable, and the two point the same way.
The device and the evidence behind it
The screws are made from a magnesium, calcium and zinc alloy (ZX00) and degrade in the body. The intended purpose covers fixation of bone fractures and fixation after osteotomies, and the listed indications run from the glenoid rim, humeral condyle, olecranon, radial head, distal radius and hand, to the ankle, talus, calcaneus, tarsals and metatarsals, together with several osteotomies and arthrodeses of the foot.
The panel found that the main clinical data was an unpublished internal report of a prospective, non-randomised feasibility study in syndesmotic injuries and dislocated fractures of the medial malleolus, conducted at a single centre. Much of the supporting literature concerned other magnesium alloys, including alloys containing rare earth elements, rather than the alloy under assessment.
What the panel found
One indication cannot stand in for many
The panel noted that the biomechanical demands differ widely between sites, observing that an olecranon fracture “differs very much from a fixation of a small fragment of the glenoid rim”. It concluded that a single feasibility study in medial malleolar fractures was insufficient to support all the listed indications, and that each indication should be described and supported by its own data. It also flagged the clinical risk: used in the wrong indication, a degradable screw could interfere with healing or allow fragments to displace.
Data from other alloys does not transfer
The benefit-risk determination rested mainly on non-clinical sources and on literature about devices that were “similar but not necessarily equivalent” to the device under assessment. Differences in alloy composition limited how far that literature could be applied.
Claims need clinical support
The claim of “rapid bone in-growth, regeneration and replacement of the implant with natural bone” was not substantiated clinically, and the panel said bioactive and osteopromotive claims would need a comparative, well-designed prospective study.
Degradation must be shown in humans
The manufacturer concluded that the implant is fully resorbed after two to three years, and that initial stability lasts six weeks, without clinical data to support either point. The panel noted that most of the screws are cannulated, which increases the surface exposed to fluid, and that mechanical strength begins to change as soon as the screw is inserted. It asked for an analysis linking degradation to fracture or osteotomy stability over time.
The PMCF plan was too thin
The panel described the information on clinical follow-up as “very scarce”, with no imaging timepoints and no defined approach to bone healing, implant degradation, gas formation or suspected osteolysis.
What the panel recommended
- A literature review covering all magnesium alloy degradable screws, including adverse events, bone healing, time to degradation and the methods used to measure it.
- A pre-authorisation clinical study across different fracture and osteotomy groups, related to the shear and compression forces at each anatomical site, with a comparison against the gold standard of fixation.
- A comparative post-authorisation study with fracture healing as the primary outcome, radiographs at six weeks, three months and one year, and CT at six months and one year to follow biodegradation. The panel noted that radiostereometric analysis could be used instead, allowing fewer patients.
- A clinical investigation within the PMCF plan to address the remaining uncertainties about safety and performance for this type of device.
The notified body is not bound by the opinion, but it must give due consideration to it and provide a full justification in its conformity assessment report wherever it does not follow the panel’s advice.
A note on timing
The opinion is dated 14 May 2024. The notified body’s request was received in March 2024, so the opinion has taken more than two years to reach publication. The Commission appears to be clearing a backlog, and it is reasonable to expect opinions from consultations some time ago to continue appearing. If your device has been through Article 54 consultation, the panel’s views on it may become public well after certification, and anything the notified body did not follow will have been justified in writing.
What this means for manufacturers
The opinion concerns one device, but the reasoning applies to any Class III implant, and particularly to degradable and novel-material implants, where the intended purpose is broader than the clinical data.
Map evidence to each indication: List every indication, site and patient group in the intended purpose and show which clinical data supports each one. Where the biomechanical or clinical demands differ, expect to show data for each group or to narrow the intended purpose. A narrower initial indication, widened as data comes in, is often the faster route to certification.
Be strict about what literature you rely on: For novel materials, small changes in composition, geometry or surface can change performance. Data on similar devices can inform the state of the art, but it will not carry the benefit-risk determination unless equivalence can be shown. Our article on literature searches and the state of the art covers how to set this up.
Tie every performance claim to clinical data: Claims about bone regeneration, resorption time, or avoiding a removal operation are exactly the claims the panel tested. If the clinical evidence does not yet support a claim, leave it out of the IFU and promotional material until it does, as the panel also required for the knee injectable.
Write a PMCF plan that answers the open questions: Specify endpoints, imaging methods and timepoints, and how degradation, gas formation and osteolysis will be measured. A PMCF plan that simply commits to collecting data will not satisfy a panel reviewing a novel implant.
Plan for Article 54 early: If your device falls within the consultation procedure, build your clinical evaluation on the assumption that an independent panel will read it. Gaps the notified body might accept may not survive panel review, and addressing them after the opinion costs far more time than addressing them before submission.
Specculo prepares and reviews clinical evaluations and PMCF plans for Class III and implantable device manufacturers, including gap reviews ahead of notified body and expert panel scrutiny. If you would like a review of how your evidence supports each indication, get in touch or read more about our regulatory consulting services.