On 18 August 2026 the US Food and Drug Administration published a discussion paper on the regulation of generative AI-enabled medical devices and asked for comments by 19 October 2026. A month later, on 17 September 2026, FDA published a final order in the Federal Register (document 2026-19074) refusing to exempt radiology computer-aided detection and triage software from premarket notification.
FDA is exploring how to evaluate generative AI devices, including whether to accept more uncertainty before market entry in exchange for stronger monitoring afterwards. At the same time, it has confirmed that an established category of AI imaging software still needs a 510(k) for every device. For EU manufacturers with US plans, both items affect how you build your evidence and your timeline.
What the discussion paper proposes
The paper comes from FDA’s Digital Health Center of Excellence. FDA states that it is for discussion purposes only and does not represent draft or final guidance. It is still the clearest indication so far of how the agency is thinking.
Risk. The paper sets out a two-axis framework. One axis is what the device does, ranging from non-directive information, through information that directs action, to taking action autonomously. The other is the consequence of relying on an incorrect output, from minor (advice on an over-the-counter remedy) to severe (adjusting an insulin dose). Risk rises as both increase. The paper also asks whether the device is patient-facing or clinician-facing, whether a generalist or specialist uses it, whether the user can check the output, whether the action is reversible, and how much time pressure there is.
Premarket evaluation. FDA describes a competency-based approach, modelled on how clinicians are assessed. Clinicians are not tested on every scenario they might meet, and the paper suggests generative AI devices could be evaluated the same way. It has two parts: device benchmarking (structured, non-clinical testing of capabilities) and clinical confirmation (evaluation in real or clinically representative settings, from retrospective review of real inputs and shadow deployment up to prospective studies). The benchmarking elements cover safety (recognising and escalating safety-critical situations, staying within scope, communicating uncertainty and deferring), clinical proficiency, generalisability across subgroups, and additional elements for agentic systems.
Comparators. On performance standards, the paper says performance “might be compared to that of a panel of qualified clinicians whose consensus reflects the applicable standard of care, or to that of a median clinician in practice.”
Postmarket monitoring. FDA writes that it “is considering whether it is appropriate to accept greater premarket uncertainty regarding a GenAI-enabled device’s benefit-risk profile through greater reliance on postmarket monitoring.” The monitoring approaches it discusses include periodic re-benchmarking, sample-based review of real-world outputs by independent clinicians, and detection of performance degradation. It also discusses predetermined change control plans (PCCPs), re-benchmarking after modifications, changes to third-party foundation models, and a voluntary master file route for foundation model developers.
The paper poses 26 questions, and respondents can answer only those relevant to them. Comments go to docket FDA-2026-N-7874 on Regulations.gov.
The CAD and CADt decision
The final order concerns a petition submitted on 22 October 2025 on behalf of Harrison.ai. It asked for a partial exemption from 510(k) for four device types: radiological computer-assisted diagnostic software for lesions suspicious of cancer, medical image analysers, radiological computer-aided triage and notification software, and radiological computer-assisted detection and diagnosis software.
The exemption would have applied to manufacturers that already held a 510(k) clearance in the relevant device type, subject to a post-market plan, transparency and training measures. FDA published notice of the petition on 29 December 2025, took comments until 27 February 2026, and denied the petition on 1 April 2026. The final order, effective on publication on 17 September 2026, concludes that the petition “does not demonstrate that premarket notification is not necessary to assure the safety and effectiveness” of these devices. Manufacturers must continue to obtain 510(k) clearance before marketing them. FDA’s detailed reasoning is in its petition response in docket FDA-2025-P-5560.
Read together, the two documents are consistent. FDA is open to leaning more on postmarket evidence for generative AI, but within premarket review. A track record in a mature AI category has not been enough to remove that step.
How this compares with the EU position
In the EU, an AI-enabled device is regulated first under the MDR or IVDR, with software qualification and classification guided by MDCG 2019-11 Rev. 1. The AI Act adds a second layer for AI systems that are safety components of, or are themselves, devices requiring notified body involvement. Following the Digital Omnibus on AI, Regulation (EU) 2026/1744, those high-risk requirements apply from 2 August 2028. We covered the delay in our post on the AI Act omnibus and medical devices.
The MDCG and the AI Board have published MDCG 2025-6, an FAQ on how the MDR, IVDR and AI Act fit together. At the time of writing there is no MDCG guidance dedicated to generative AI. EU manufacturers therefore apply the general requirements (clinical evaluation or performance evaluation, risk management, software lifecycle, post-market surveillance) to generative AI without a specific framework.
There is overlap in substance. Both systems expect risk to follow intended use and the consequence of error, and both place weight on post-market surveillance. The main practical difference is change management. FDA has an established PCCP mechanism for AI-enabled device software, while under the MDR a significant change to a certified device goes through the notified body. The AI Act allows for changes that the provider has predetermined at the initial conformity assessment, which will matter once it applies to devices.
What to do next if you plan to enter the US
Confirm your US pathway early. Identify the product code, classification regulation and likely predicate. If your device falls under one of the four CAD or CADt regulations, plan for a 510(k) for each new device, including where you already hold a clearance.
Read the discussion paper against your own evidence. If your device uses generative AI, map your testing to FDA’s benchmarking elements and consider which clinician comparator you would use. If you have a view on any of the 26 questions, the deadline is 19 October 2026.
Build one evidence base for both markets. A benchmarking study against a defined clinician comparator, with subgroup analysis, can support both an MDR clinical evaluation or IVDR performance evaluation and a US submission if it is designed with both in mind from the start.
Design post-market monitoring that serves both regulators. Re-benchmarking at set intervals, sampled clinician review and drift detection fit naturally into an MDR post-market surveillance plan and are the kind of monitoring FDA describes.
Plan change control for each jurisdiction. Decide what belongs in an FDA PCCP, and how the same changes will be handled under your notified body’s significant change process.
Control your foundation model supply. If you build on a third-party model, make sure your supplier agreements give you notice of model changes and enough information to assess them.
Use a Pre-Submission. For a novel generative AI device, FDA’s Q-Submission programme is the practical way to test your evidence plan before you commit to it.
For the EU side of this work, our SaMD developer guide to the MDR and IVDR and our overview of AI regulation are good starting points.
Specculo advises SaMD and AI device manufacturers on regulatory strategy, including evidence plans designed to serve both EU certification and a later US submission. If you are planning US entry for an AI-enabled device, get in touch and we can talk it through.
Sources
- FDA: Considerations for the Regulation of Generative AI-Enabled Medical Devices, discussion paper and request for feedback
- FDA discussion paper (PDF)
- Federal Register: Medical Devices; Exemption From Premarket Notification: Radiology Computer-Aided Detection and/or Diagnosis Devices and Computer-Aided Triage and Notification Devices (FR Doc. 2026-19074)
- Regulation (EU) 2026/1744 (Digital Omnibus on AI), EUR-Lex
- MDCG 2025-6: FAQ on the interplay between the MDR, IVDR and the AI Act (June 2025)