Notified Bodies Get a Stopwatch: What Commission Implementing Regulation (EU) 2026/977 Means for Manufacturers

On 5 May 2026, the European Commission published Implementing Regulation (EU) 2026/977, laying down uniform quality management and procedural requirements for notified bodies under both the Medical Devices Regulation 2017/745 and the In Vitro Diagnostic Regulation 2017/746. Adopted on 4 May, it is one of the most consequential pieces of secondary legislation we have seen in this space since the regulations themselves, and it directly addresses three areas that have caused industry pain for years: predictability of timelines, transparency of quotations, and consistency of re-certification.

Below is what the Regulation actually requires, what changes from current practice, and how the new framework should shape your planning.

Why This Regulation Exists

The recitals do not pull punches. The Commission notes that notified bodies have applied diverging practices on quotations, on how timelines for conformity assessment activities are agreed and managed, and on how re-certification is scoped. The result has been wide variation in cost and turnaround across notified bodies for materially similar work, with disproportionate impact on small and medium-sized manufacturers and on innovation more broadly.

Article 36(3) MDR and Article 32(3) IVDR give the Commission the power to specify uniform quality management and procedural requirements where the existing Annex VII text is being applied inconsistently. Implementing Regulation 2026/977 is the exercise of that power. The legal effect is that every designated notified body in the EU will, from the application date, work to the same procedural standards on these matters.

The Maximum Timelines

This is the part of the Regulation that will get the most attention, and rightly so. Article 2 sets binding maximum timelines for the core conformity assessment activities. Notified bodies remain obliged to agree the shortest feasible timeline with the manufacturer, taking into account the device type, technology, risk class and conformity assessment route. The maxima are the ceiling, not the target.

For an initial conformity assessment, the limits are:

  • 30 days for the application review and the signature of the contract, running from receipt of a complete application to contract signature.
  • 120 days for quality management system auditing under Section 4.5.2 of Annex VII, running from the audit programme’s first activity to the close of the final review.
  • 90 days for product verification under Section 4.5.3, running from the start of technical documentation assessment to the close of the final review.
  • 20 days for the decision and certification, running from the close of the final review to the entry of the certificate in Eudamed.

Where the procedure permits parallel running of QMS audit and product verification (Annex IX MDR or IVDR), the Regulation explicitly confirms this is allowed, provided the audit programme reflects relevant input from the technical documentation assessment.

For substantial changes to the QMS, to the device range covered by an EU QMS or quality assurance certificate, or to an approved device covered by an EU technical documentation assessment or type-examination certificate, the limits are:

  • 30 days for the notified body to assess the proposed change and decide whether further conformity assessment activities are needed.
  • 90 days for any additional conformity assessment activities required.
  • 20 days for issuing the supplement to the affected certificate.

For re-certification (covered separately in Articles 5 to 7):

  • 90 days maximum for the notified body’s assessment of the renewal application and the supporting documentation.
  • 20 days for the decision and re-issuance of the certificate, with a derogation if the decision is taken more than three months before the certificate’s expiry date (in which case the 20 days runs from the three-month mark).

These are real ceilings. The Regulation makes clear that hitting the ceiling, or running out of permitted interruptions (more on which in a moment), is not by itself a basis for the notified body to refuse a certificate. The conformity assessment must continue until a positive or negative decision is made.

Interruptions

The maximum timelines come with one major caveat that deserves its own section. Notified bodies may interrupt the clock when they are waiting for the manufacturer to address non-compliances, or to answer duly justified questions and requests necessary for the assessment. Article 3 caps how many times this can happen.

For an initial assessment, the limits are:

  • One interruption during the 30-day application review and contract phase.
  • Four interruptions during QMS auditing, with two additional interruptions permitted for each additional site that needs to be audited on-site.
  • Four interruptions during product verification.
  • One interruption during the 20-day decision and certification phase, where the notified body asks the manufacturer to verify the certificate information and, if needed, complete Eudamed entry.

For substantial change phases (a) and (b) combined, the cap is five interruptions in total. For the reviews and verifications under Articles 5 and 6 (re-certification), the cap is three interruptions in total, plus one for the certificate issuance step.

Where notified body and manufacturer agree on a rolling review of technical documentation, they may also agree additional interruptions on top of the figures above, with a plan for how parts of the documentation are submitted.

Crucially, interruptions while waiting for an EMA opinion, a regulatory authority, an expert panel or an EU reference laboratory contribution are not counted in the figures above. Those external dependencies sit outside the cap.

The practical implication is that every clarification round is now a finite resource. Manufacturers who reply piecemeal, or who treat each notified body question as an isolated request rather than as part of a managed response, will exhaust their allowance and the timeline will simply continue. Structuring responses, escalating questions internally to the right subject-matter expert quickly, and consolidating non-compliance closure into clear submissions becomes a planning discipline rather than a nice-to-have.

What Changes About Quotations

Article 1 standardises the front end of the engagement. Notified bodies must operate documented procedures so that quotations are only issued after they have received from the manufacturer a defined set of inputs, including:

  • Manufacturer identification, including the information needed to determine whether the manufacturer is a micro, small or medium-sized enterprise under Recommendation 2003/361/EC (number of employees and annual turnover).
  • The authorised representative, where applicable.
  • Each site covered by the QMS, with addresses, headcount, shifts, and the activities performed.
  • Suppliers and sub-contractors that perform design and manufacturing activities relevant to the conformity assessment, with a description of those activities.
  • A description of the device, intended purpose, specific characteristics or technologies used, and the risk class.
  • The conformity assessment procedure being applied for.
  • For changes and re-certifications, a description of the planned changes and identification of the affected certificates.

For change and re-certification quotations, the notified body must refrain from re-requesting the standing manufacturer and device information if the manufacturer confirms it has not changed. That removes a source of unnecessary friction in repeat engagements.

The quotation itself must include estimated overall costs broken down between QMS assessment and technical documentation assessment, typical costs for surveillance activities and unannounced audits, an estimation of potential extra costs (with hourly fees only acceptable where the duration of an activity cannot reasonably be predetermined), and the estimated timeline.

There is also a real obligation around cost movement. Any increase above 10 per cent of the estimated cost must be flagged to the manufacturer in advance, with a reason. That ends the experience of opening an invoice substantially higher than the original quote with no advance warning.

The recitals encourage notified bodies to use structured pre-application dialogue with manufacturers to scope the work properly before quoting. The intention is clear: better information up-front, better quotes, fewer surprises.

Re-Certification, Reframed

Articles 5 and 6 reset the scope of re-certification, an area where notified body practices have varied so widely that some renewals look like targeted reviews and others look indistinguishable from a fresh certification. The Regulation pushes hard towards focused, evidence-based re-certification.

For product certificates, manufacturers must provide a defined renewal package: a list of changes (notified or not) since the last certification, the most recent periodic safety update report and a summary of field safety corrective actions, a summary of changes to the risk evaluation that affected the benefit-risk ratio, identification of changes to keep pace with state of the art, the most recent clinical evaluation or performance evaluation report, and identification of any other changes notified under Section 4.11.

Manufacturers must also provide a list of changes that are not yet notified but are needed to keep the device aligned with new regulatory requirements, common specifications, scientific findings and standards. Where the basis is new scientific findings, the source category has to be indicated (new medical or technical knowledge, new test methods, findings on materials, or clinical / performance evaluation results from equivalent devices).

Notified bodies have 90 days from receipt of that documentation to assess it. The assessment is structured: verify coherence with post-market surveillance, with state of the art, with the updated risk analysis; confirm non-compliances are resolved or under accepted CAPA; check certificate conditions and limitations; check whether the scope needs amending; complete the final review.

For QMS re-certification, the 90-day clock runs from receipt of the renewal application. The notified body must verify that all relevant audit requirements have been fully assessed at least once during the certification cycle, that surveillance results still demonstrate compliance, that audit programme and sampling plans remain appropriate, that non-compliances are resolved or under CAPA, and that scope, conditions and limitations remain appropriate.

Article 7 sets the decision and re-issuance limit at 20 days from the close of the final review. Where the decision is taken more than three months before expiry of the existing certificate, the 20-day clock starts at the three-month mark. The Regulation also makes clear that re-certification is to be limited to assessing the documentation set out in Articles 5 and 6, with additional information requests permitted only where strictly necessary to complete the assessment. Re-certification should not become a second initial certification.

Monitoring and Public Reporting

Article 4 introduces the transparency mechanism that, over time, may matter as much as any of the headline timelines. Notified bodies must establish a monitoring system within their QMS that captures, for each completed conformity assessment activity, the percentage completed within the maximum timelines, the median duration from application to certification in days, and the median total cost in euros. This applies to initial assessments under Annexes IX, X and XI of MDR and IVDR, and to assessments of changes under Article 2(3) of the new Regulation.

By 30 April each year, every notified body must draw up an annual report covering the previous year’s activities, publish it on its website, and inform its designating authority and the Commission. Article 4(4), the publication obligation, applies from 1 January 2028.

For manufacturers, that means publicly comparable performance data on every notified body in the EU from 2028 onwards. For the first time, the choice between notified bodies will be informed by something more than reputation and market gossip. That comparability is likely to drive its own behavioural shift, ahead of any direct enforcement of the timelines themselves.

When It Applies and What It Does Not Touch

The Regulation enters into force on 25 May 2026, twenty days after publication in the Official Journal. Application is staggered:

  • Articles 1, 2 and 3 (quotations, timelines, interruptions) apply from 25 February 2027, and do not apply to conformity assessment procedures for which the notified body and manufacturer signed a written agreement before that date.
  • Article 4(1) to (3) (the monitoring system itself) applies to procedures with contracts signed after 25 May 2027.
  • Article 4(4) (the annual public report) applies from 1 January 2028.
  • Articles 5, 6 and 7 (re-certification) do not apply to re-certification reviews of certificates expiring before 25 November 2027.

In short, current files in flight are unaffected. The new framework reshapes how applications signed in 2027 onwards will be planned and priced, and how re-certifications of certificates expiring in late 2027 and beyond will be scoped.

What Manufacturers Should Do Now

The Regulation is a positive development for the industry, but the benefits will accrue mainly to manufacturers who plan ahead of the 25 February 2027 application date rather than reactively. A few practical actions are worth taking in the next twelve months.

Map your certificate calendar against the application dates. Any contract signed before 25 February 2027 falls outside Articles 1 to 3. Re-certification reviews of certificates expiring before 25 November 2027 fall outside Articles 5 to 7. Knowing where your portfolio sits relative to those cut-offs tells you which engagements you can plan under the new framework and which will run under existing notified body practice.

Look at your re-certification readiness. The Article 5 documentation set is more focused than some notified bodies have demanded in the past, but it still requires a reasonably mature picture of post-market surveillance, state of the art tracking, risk evaluation updates, and clinical or performance evaluation refresh. If those processes are running well, your re-certification will be smooth. If they are patchy, the 90-day notified body window is short, and the Article 5(5) mechanism allowing additional information requests is bounded to what is “necessary” rather than open-ended.

Watch for notified body procedural updates. Every designated notified body in the EU will need to update its documented procedures to align with the Regulation. The pace and quality of those updates will vary. From 2028, you will be able to compare notified body timelines and costs in public reports. Until then, ask your notified body what their implementation roadmap looks like.

Need Help Working Out What This Means for Your Files?

The new framework is straightforward in principle but the practical implications are different for every manufacturer, depending on certificate calendar, conformity assessment route, notified body, and the maturity of your internal post-market and change-control processes. If you’d like to walk through what 2026/977 means specifically for your files, your renewal pipeline, or your next notified body engagement, we can help.

Book a call with Kenneth and we’ll work through the impact together.

Source: Commission Implementing Regulation (EU) 2026/977 (Official Journal)