Guidance on Key Updates Introduced to ISO 10993-1:2025

Introduction

Edition 6 of ISO 10993-1 (ISO 10993-1:2025) was published in November 2025 and introduces significant changes to previous editions. This article intends to support medical device manufacturers in understanding and implementing the key changes introduced in the revised edition of ISO 10993-1. The following sections summarise the principal updates, clarifications, and new expectations introduced in this edition of the standard. 

1) Integration with ISO 14971 Risk Management

The biological evaluation process is now fully embedded within the ISO 14971 risk management framework. Biological hazards must be identified, evaluated, controlled, and reviewed in the same manner as other product risks.

The Biological Evaluation Plan, biological risk assessment activities, and Biological Evaluation Report must be clearly traceable to the risk management file. This reinforces the expectation that biological safety is not a standalone exercise, but an integral part of the overall assessment of device safety and performance.

In practice, this means that conclusions drawn in the biological evaluation must align with risk management decisions, including risk control measures and benefit-risk determinations. Standalone biological reports that are disconnected from the risk management process are increasingly unlikely to be considered acceptable.

2) Consideration of Reasonably Foreseeable Misuse

A significant update in the revised ISO 10993-1 is the explicit requirement to consider reasonably foreseeable misuse as part of the biological evaluation. Manufacturers must now assess not only the intended use of a medical device, but also predictable patterns of misuse that could reasonably occur and that may alter biological exposure or risk.

This does not require consideration of every conceivable misuse scenario. Rather, foreseeable misuse must be supported by objective evidence such as published literature, post-market surveillance data, complaint trends, or recurring documented patterns observed in similar devices. Any identified misuse scenarios should be systematic and repeatable rather than isolated or one-off cases.

Where foreseeable misuse is identified, its potential impact on biological safety must be addressed within the overall biological risk assessment.

3) Transition from “Endpoints” to “Biological Effects”

In parallel with the simplified contact categorisation, the revised standard no longer uses the term “endpoints”. Instead, the concept of “biological effects” is applied throughout the document.

This change reflects a broader shift away from a purely test-driven checklist approach towards a more holistic evaluation of how device materials, constituents, and degradation products may interact with the human body. Manufacturers are expected to consider the full range of relevant biological effects based on scientific understanding and exposure, rather than focusing solely on predefined test endpoints.

4) Simplified Contact Categories

The revised ISO 10993-1 simplifies the classification of medical devices according to the nature of body contact. Devices are now categorised into four contact groups:

  1. Devices in contact with intact skin.
  2. Devices in contact with intact mucosal membranes.
  3. Devices in contact with breached or compromised surfaces or internal tissues other than blood.
  4. Devices in contact with blood.

Externally communicating medical devices are no longer identified as a separate contact category. Biological effects for such devices are now assessed based on the specific tissues or fluids they contact.

  • Devices that interface with internal tissues, bone, dentin, or breached or compromised skin are evaluated under the category of contact with breached or compromised surfaces or internal tissues other than blood. 
  • Devices that have direct or indirect contact with blood, including those that store blood or act as conduits for fluids to or from the bloodstream, are evaluated under the blood-contacting category. 

This approach places greater emphasis on accurately defining the true nature of body contact when determining applicable biological effects.

5) Updates to Biological Effect Evaluation

Systemic Toxicity Evaluation

The previous edition of ISO 10993-1 differentiated between acute, subacute, subchronic, and chronic systemic toxicity. In the revised edition, these distinctions have been consolidated into a single category referred to as systemic toxicity.

The standard clarifies through explanatory notes that evaluations of systemic toxicity must still be appropriate to the duration and nature of device contact. The consolidation is intended to simplify terminology, not to reduce the scientific rigour of systemic toxicity assessments.

Material-Mediated Pyrogenicity

Material-mediated pyrogenicity has been removed as a routine biological effect requiring testing. This change reflects evidence that material-mediated pyrogenic responses are rare and typically associated with a limited number of specific substances.

Guidance on evaluating material-mediated pyrogenicity is now included within the section addressing other biological effects. Where relevant, assessment should be risk-based and supported by chemical characterisation and toxicological rationale rather than default testing.

Genotoxicity Considerations

Under the revised ISO 10993-1, genotoxicity evaluation is now applicable to all medical devices with prolonged contact, with the exception of devices that contact intact skin only.

Manufacturers must therefore ensure that genotoxic potential is appropriately addressed for prolonged contact devices, either through testing or through robust scientific justification based on material composition, chemical characterisation, and exposure assessment.

Local Effects After Tissue Contact

The term “implantation effects” has been replaced with “local effects after tissue contact”. This change clarifies that local biological effects are not limited to implanted devices.

Manufacturers are expected to consider local tissue responses for any device that contacts internal tissues, including irritation, inflammation, or other local effects that may arise from transient, repeated, or long-term contact.

6) Contact Duration and Bioaccumulation

The revised standard clarifies that contact duration is determined by the total number of days on which a medical device or its components come into contact with tissues or circulating blood, irrespective of the length of contact on any given day. In addition, manufacturers must now consider the potential for bioaccumulation when assessing contact duration. Where materials contain substances known to bioaccumulate in humans, the device should generally be treated as having long-term contact unless a robust scientific justification supports an alternative classification. 

7) Introduction of Biological Equivalence

The revised standard introduces and formalises the concept of biological equivalence. Biological equivalence may be claimed when the chemical and physical characteristics and manufacturing processes of a medical device or material under evaluation are sufficiently similar to those of a comparator device or material with established biological safety. 

Criteria used to demonstrate biological equivalence may include:

  • Type and duration of body contact
  • Material composition and formulation
  • Relevant physical or mechanical properties
  • Manufacturing processes, including sterilisation and packaging
  • Intended patient population
  • Quantity or size of material

Any differences between the device and the comparator must be demonstrated not to introduce new or increased biological risks.

8) Gap Analysis as a Core Element of Biological Evaluation

The revised standard introduces a clear expectation for the performance of a gap analysis early in the biological evaluation process. A gap analysis is a structured review of the information collected to determine whether it is sufficient to support a biological risk assessment in accordance with ISO 10993-1. It is conducted immediately after the initial data gathering and is intended to determine whether the available information is adequate to demonstrate biological equivalence or to proceed directly to biological risk estimation. 

If the gap analysis identifies insufficient information, additional data must be obtained or generated. This may involve: 

  •  further physical or chemical characterisation of materials;  
  • review of relevant clinical data; or  
  • biological testing of the finished medical device.  

The gap analysis must be repeated after additional data are collected to confirm that remaining uncertainties have been adequately addressed. All gap analyses, findings, and subsequent actions must be fully documented. 

9) Biological Evaluation Documentation Requirements

The revised ISO 10993-1 expands and clarifies the required content of both the Biological Evaluation Plan and the Biological Evaluation Report. Documentation must clearly demonstrate how biological hazards were identified, evaluated, and controlled, and how these activities integrate with the overall risk management process.

Clear traceability between biological evaluation activities, supporting evidence, and risk management conclusions is essential.

10) Lifecycle-Based Biological Evaluation

Finally, the revised ISO 10993-1 places strong emphasis on lifecycle-based biological evaluation. Manufacturers are expected to consider biological safety from material selection and manufacturing through clinical use and post-market surveillance. Biological risks must be reassessed as new information becomes available to ensure continued safety throughout the device lifecycle. 

Conclusion

The revisions introduced in ISO 10993-1:2025 reflect a continued shift away from prescriptive, test-driven biological evaluation towards a more integrated, risk-based, and lifecycle-oriented approach. While many of the underlying scientific principles remain unchanged, the standard now places greater emphasis on justification, traceability, and the quality of the manufacturer’s rationale.

For manufacturers, this means that legacy biological evaluations may no longer be sufficient without review. Areas such as foreseeable misuse, cumulative contact duration, bioaccumulation, biological equivalence, and formalised gap analysis are likely to attract closer scrutiny during conformity assessment. In practice, the revised standard raises expectations around documentation quality and consistency with risk management, rather than simply introducing new testing requirements. But with that said, the concepts such as biological equivalence and gap analysis allow manufacturers to avoid conducting additional testing where not necessary.

Early engagement with the updated requirements can help avoid delays during regulatory review and ensure that biological evaluations remain robust, defensible, and proportionate to device risk throughout the product lifecycle.

How Specculo Can Support ISO 10993-1:2025 Implementation

Adapting to ISO 10993-1:2025 often requires more than incremental updates to existing documentation. In many cases, manufacturers need a structured reassessment of their current biological evaluation strategy to confirm alignment with the revised expectations.

Specculo supports manufacturers by performing targeted gap assessments against ISO 10993-1:2025, focusing on identifying where existing Biological Evaluation Plans, Biological Evaluation Reports, and supporting evidence may fall short of the new requirements

Where gaps are identified, our team can support the development or revision of Biological Evaluation Plans and Biological Evaluation Reports in line with the 2025 edition of the standard. This includes helping manufacturers structure defensible rationales, align biological evaluation outputs with risk management documentation, and ensure that evaluations are suitable for review by notified bodies and other regulatory stakeholders.

This support can be provided either as a standalone activity or as part of a broader regulatory or technical documentation engagement, depending on the manufacturer’s needs and internal capabilities.