ISO 10993-1 Final Draft: Key Biocompatibility Changes

A draft of the new revision of ISO 10993-1, addressing the biological evaluation of medical devices, was released at the end of June 2024, containing a number of fundamental updates. The final draft has just been published. These potential updates introduce a significant shift in the way medical device biocompatibility is evaluated, reflecting an evolution toward a more dynamic, risk-based approach. These updates bring essential implications for manufacturers, regulatory professionals, and testing laboratories involved in the biological evaluation of medical devices. 

Below is a breakdown of the critical updates and their practical consequences.

1. Lifecycle-Based Biocompatibility Evaluation 

The revised draft places a strong emphasis on evaluating biocompatibility throughout the device’s lifecycle; from manufacturing to shelf-life to use and disposal. While the previous version of the standard acknowledged lifecycle considerations, the revised draft integrates this concept throughout the document. Manufacturers would now be required to: 

  • Justify that biocompatibility remains stable throughout the product lifecycle. 
  • Demonstrate that no harmful substances are released over time.
  • Ensure tested samples are representative of the device at different stages, including at product expiration. 

2. Consideration of Foreseeable Misuse 

The revised draft standard explicitly requires evaluating biocompatibility risks associated with foreseeable misuse. Examples include: 

  • Devices used longer than intended. 
  • Use by unintended patient populations, such as pediatric patients using adult-only devices. 

This shift emphasises the importance of accounting for user behaviour and off-label use during the risk assessment process. 

3. Greater Focus on Material Characterisation 

More detailed material data will need to be gathered or generated, especially for complex or novel materials 

4.  Expanded Need for Genotoxicity Testing 

The standard now acknowledges that genotoxicity risks may be more prevalent than previously assumed, particularly in cases with potential systemic toxicity. While genotoxicity tests may still be unnecessary for devices contacting only intact skin, more devices will now require this evaluation as part of a comprehensive risk-based assessment.Shape 

5. Introduction of the Device Family Concept 

A new and practical concept that could potentially be introduced is the Medical Device Family. This approach allows manufacturers to: 

  • Group devices with the same materials and manufacturing processes. 
  • Perform biocompatibility testing on a worst-case device to represent the entire family. 

This offers a more efficient and strategic path for testing while maintaining robust safety standards. 

6. Qualification Requirements for Evaluators 

The draft emphasises that biological evaluations must be performed by qualified professionals. The evaluators’ education, training, and experience must be documented, and CVs must be submitted to validate competency. 

7. Structured BEP (Biological Evaluation Plan) Requirements 

A major procedural update is the mandatory documentation of the Biological Evaluation Plan (BEP). Now supported by a new flowchart, the BEP must include: 

  • Specific acceptability criteria for biological risks. 
  • Test strategies and their rationale. 
  • Integration with ISO 14971-based risk assessments. 

This replaces the vague guidance of previous versions with a clear framework for planning and documenting biocompatibility assessments. 

8. Redefinition of Contact Duration for Multiple Exposures 

For devices with multiple intended exposures, the contact duration category must be based on the total exposure period for a single patient, not the cumulative contact time. For example: 

  • A device used for 10 minutes daily over 7 days is considered 7-day contact, not 70 minutes. 
  • “Very brief” contact (<1 minute) may still warrant justification. 
  • Any contact >1 minute defaults to 24-hour exposure for classification purposes. 

9.  Simplified Contact Categories 

The number of contact categories has been reduced to 4. 

10. Reduced Need for Material-Mediated Pyrogenicity Testing 

The revision introduces a reduction in the requirement for testing material-mediated pyrogenicity. 

11. New Guidance on Animal Welfare 

A dedicated section reinforces the ethical consideration of animal welfare, in line with global trends to reduce, refine, and replace animal use (in vivo) in testing. 

Conclusion 

The final draft of ISO 10993-1 marks a pivotal change in the biological evaluation of medical devices. By embedding a lifecycle and risk-based approach, the standard aims to improve patient safety. Biological safety is not just about passing tests, but about demonstrating long-term safety through scientific rigour, risk management, and lifecycle thinking. 

Medical device developers and regulatory teams should start aligning their processes now in anticipation of the finalised standard. Staying ahead of these changes will be key to efficient submissions and market access in the coming years.