On 9 September 2026 the Medical Device Coordination Group published MDCG 2020-16 rev.5, the latest revision of its guidance on the IVDR classification rules. The revision history lists a single change: Rule 7, with a “clarification of the rationale and revision of examples”. No other section is listed as changed.
A standalone quality control material can give the presence or concentration range of the analyte it contains and still be class B under Rule 7, because those figures are not treated as assigned values. That outcome depends on how the instructions for use describe the product’s intended purpose. Controls with values assigned by the manufacturer, including assay kit validity controls, continue to take the class of the device they control.
What Rule 7 says now
Rule 7 itself is unchanged: “Devices which are controls without a quantitative or qualitative assigned value are classified as class B”. The revision rewrites the rationale beneath it.
The guidance explains that diagnostic laboratories must continuously monitor the validity of their results, and that standalone QC materials may be used for this “alongside the control material supplied by reagent or instrument manufacturers”. It then sets a condition on labelling: “It must be clearly stated in the IFU that the intended purpose of these standalone QC materials is limited to monitoring the validity of laboratory examination procedures. They cannot replace the mandatory QC materials required to verify the performance of the IVD itself.”
On analyte information, the text is more direct. Standalone QC materials need properties that suit the examination method being monitored, so “information on the presence or concentration range of the analyte contained in the QC material may be necessary and can be provided by the manufacturer. Such values indicated by the manufacturer of the standalone QC materials are not considered ‘assigned values’. Unless other rules apply, this type of control is classified as class B according to rule 7.”
The rationale also confirms that a manufacturer may indicate a range of values for each parameter in the IFU, while the target values the laboratory uses routinely are the ones the laboratory determines itself.
Where the line sits
The revised rationale draws the boundary in one sentence: “Controls used in the context of diagnostics that have quantitative or qualitative values assigned by the manufacturer (including assay kit validity controls) are classified according to implementing rule 1.6.” Under implementing rule 1.6, those controls fall into the same class as the device they are intended to control, so they can be class B, C or D.
The guidance also repeats that internationally certified reference materials and materials used for external quality assessment schemes are not IVDs, citing Article 1(3)(c) and (d) of the IVDR.
The revised examples of Rule 7 devices include unassigned control sera, unassigned heterozygous quality controls for nucleic acid extraction, amplification and detection, non-assay specific control plasma for coagulation and non-assay specific control serum containing multiple biochemical analytes.
What this means for manufacturers
If you manufacture third-party or independent QC materials, the IFU now carries much of the classification argument. We would check three things:
First, read your intended purpose statement against the guidance wording. It should say clearly that the product is limited to monitoring the validity of laboratory examination procedures. Our view is that the IFU should also say, in plain terms, that the material does not replace the QC material required to verify the performance of the IVD itself, since that is the distinction the MDCG relies on.
Second, look at how analyte information is presented. Indicative presence or concentration ranges are acceptable. Anything that reads as a value assigned by you for a specific assay moves the product towards implementing rule 1.6 and a potentially higher class.
Third, if your classification rationale in the technical documentation relied on the earlier wording of the guidance, update it to cite rev.5 and the revised examples. Where you had classified a standalone QC material above class B because its IFU quotes analyte ranges, it is worth revisiting that decision with your notified body.
Take care before changing IFU wording on a device that is still on the market under the IVDR transitional provisions. Those devices must not undergo significant changes in design or intended purpose, so assess any wording change against the significant change criteria first. For class B legacy devices, the formal application to a notified body is due by 26 May 2027 and the written agreement by 26 September 2027.
If you would like a second view on the classification rationale or IFU wording for a QC material, get in touch. For background on how the IVDR is structured, our overview of how IVDs are regulated in the EU is a useful starting point, and our technical file consulting team works on classification rationales regularly.