The MDCG has published a revision of MDCG 2021-24, the guidance on classification of medical devices under MDR Annex VIII. Rev.1 was released in April 2026 and updates the October 2021 version that most manufacturers have been working from.
The classification tables have new examples, and one definitional change in particular could affect how you approach classification for certain device types.
Here is what you need to know.
The definitional change that matters most
Section 3.1.4 covers invasiveness in the context of implantable devices. In the 2021 version, the definition referenced “surgical procedure.” Rev.1 replaces that with “clinical procedure,” and goes further by adding explicit clarification: “The term ‘clinical procedure’ will encompass both surgical and non-surgical procedures.”
This is a quiet but meaningful shift. Devices implanted through non-surgical routes were already subject to classification rules, but the guidance language now makes that explicit. If your classification rationale leans on a reading of “surgical” as the relevant threshold, it is worth reviewing against the revised definition.
Rule 8: two new practical notes
Rule 8 covers implantable devices and long-term surgically invasive devices, both active and non-active. In the 2021 guidance it had five practical notes. Rev.1 adds two new ones, bringing the total to seven.
The first new note (Note 1) addresses the Article 52(4) exemption. This exemption allows sutures, staples, dental fillings, braces, crowns, screws, wedges, plates, wires, pins, clips, and connectors to be assessed as a group for the purposes of notified body technical documentation review, rather than on a per-device basis. The new note clarifies that this exemption does not automatically make all such devices Class IIb. They must still be classified individually under the applicable rules. The exemption is about the conformity assessment route, not about classification outcome.
The second new note (Note 7) deals with spinal column contact. It confirms that contact with the spinal column means contact with any bony structure of the column, covering cervical, thoracic, lumbar, sacral, and coccyx regions, including spinous and transverse processes. It also confirms that hooks used to fix rods on the spinal column fall within the same treatment as screws, wedges, plates, and instruments already referenced in the Rule 8 practical notes.
Both additions are clarifications rather than new policy, but they resolve ambiguities that have generated inconsistent classification decisions in practice.
Updated classification tables: Rules 10 and 12
Rules 10 and 12 have both had their classification tables updated with new examples.
Under Rule 10, which covers active devices for diagnosis and monitoring, the Class I examples now include surgical lamps. The Class IIa examples have been expanded to include devices for visual electrophysiology and fundoscopes for examination of the eye. These additions give clearer reference points for diagnostic devices that may have fallen into a grey area in the 2021 guidance.
Under Rule 12, which covers active devices intended to administer or remove medicinal products and other substances to or from the body, the Class IIa table now includes elastomeric pumps and balloon pumps for infusion. The pressure regulators entry has also been refined to “pressure regulators for medical gases with integrated flowmeter,” making the scope of that example more precise.
If any of your devices sit in these categories, review your classification rationale against the updated tables.
Rule 9: structural reorganisation
Rule 9 covers active therapeutic devices that administer or exchange energy, as well as devices that control or influence active implantable devices. In the 2021 version, Note 1 under Rule 9’s practical issues contained guidance on illumination devices, specifically stating that devices used for illumination in the process of diagnosis (visual electrophysiology, fundoscopes for eye examination) should be Class IIa or higher under Rule 9, while devices used for illumination for therapeutic purposes are Class I.
In Rev.1, that language has been removed from Rule 9’s notes and those devices have been added directly to the Rule 10 classification table as explicit Class IIa examples. The change was described as “structure changed to better match the rule.” The practical effect is that the classification of these diagnostic illumination devices is now anchored in Rule 10 rather than Rule 9.
Rule 16: steriliser scope clarified
Rule 16 covers devices specifically intended for disinfecting, cleaning, rinsing, hydrating, or sterilising. The Class IIa sterilisers example was updated in Rev.1. The original read: “Sterilisers intended to sterilise medical devices in a medical environment.” Rev.1 reads: “Sterilisers intended to sterilise medical devices (invasive or non-invasive) in a medical environment.” The addition of “(invasive or non-invasive)” clarifies that Class IIa covers sterilisers for both device types when used in a medical environment, distinguishing them from the Class IIb row which covers washer-disinfectors intended for disinfecting invasive devices as the end point of processing.
Rule 22: AED pads and electrodes clarified
Rule 22 covers active therapeutic devices with an incorporated diagnostic function: closed-loop systems, automated external defibrillators, automated insulin delivery systems, deep brain stimulation devices, and similar.
The table change in Rev.1 is specific to AEDs. The original example read “Automated external defibrillators (AED) including their pads/electrodes.” Rev.1 narrows this to “including their pads/electrodes if they are an integral part of the defibrillators.” The new practical Note 2 then addresses the standalone case: “Defibrillators’ pads/electrodes placed on the market on their own may be considered under Rule 9 or Rule 1.” This is a meaningful clarification for manufacturers of AED accessories sold separately from the defibrillator unit itself.
The “device” to “product” change in section 3.2
Section 3.2 covers the application of classification rules. Rev.1 replaces “device” with “product” throughout this section. This may seem like a minor editorial point, but if your technical documentation quotes directly from section 3.2, the language will no longer match the current guidance. It is worth updating any direct quotations to maintain consistency.
What you should do now
If you have existing technical documentation that references MDCG 2021-24, the first step is to identify where the classification justification sits and which rules it relies on.
For implantable devices and long-term surgically invasive devices (Rule 8), check whether your rationale touches the Article 52(4) exemption or spinal column contact. If it does, verify that your classification conclusion and the reasoning behind it align with what the new notes clarify.
For active diagnostic and monitoring devices (Rule 10) or infusion-related active devices (Rule 12), compare your device against the updated classification tables. If new examples have been added in a class you are already in, your classification is likely well-supported. If the examples suggest a different class, that warrants a closer look.
For implantable devices, review your use of the invasiveness concept in section 3.1.4. If your classification rationale draws on the surgical/non-surgical distinction, update the language to reflect the revised definition.
For any documentation that quotes section 3.2 directly, update the quotations from “device” to “product.”
We work with manufacturers across device classes and risk categories to keep technical documentation current with evolving guidance. If you want to talk through how MDCG 2021-24 Rev.1 applies to your device, get in touch.