In July 2026 the Medical Device Coordination Group published MDCG 2026-5, a short position paper on who is allowed to assign a Unique Device Identifier. The answer is the manufacturer, and only the manufacturer. If a distributor sells your device under its own brand, that does not change who owns the UDI.
The short version is this. Two UDI-DIs for the same device sold under two trade names can be correct. But both must be assigned by the manufacturer, both must be linked to the manufacturer at the issuing entity, and only the manufacturer can register the device in EUDAMED. A distributor cannot obtain a UDI-DI code in its own name for a device it did not manufacture.
What prompted the paper
The MDCG had seen a practice worth calling out. Under Article 16(1)(a) of the MDR and IVDR, a distributor can make a device available on the EU market under its own name while keeping the manufacturer’s details on the label. That part is legitimate. The problem was what some companies did following that.
In the arrangement the MDCG describes, the distributor goes to an EU UDI issuing entity and obtains a UDI-DI code in its own name. The device then ends up with two groups of UDI-DIs registered under the same Basic UDI-DI: those the manufacturer assigned for its own brand, and those the distributor assigned for the same device under the distributor’s brand. In EUDAMED, the same device carries identifiers linked to two different economic actors. That is the situation MDCG 2026-5 sets out to correct.
What the regulation actually says
The paper anchors its position in the text. Annex VI, Part C of both the MDR and the IVDR states that the manufacturer shall assign and maintain unique UDIs for its devices, and that only the manufacturer may place the UDI on the device or its packaging. The UDI-DI is defined as “specific to a manufacturer and a device”. Registration obligations under Articles 29 and 31 MDR, and 26 and 28 IVDR, sit with the manufacturer.
Read together, the MDCG’s conclusion is that no other economic actor, whether an authorised representative, distributor or importer, may assign UDIs to a device. The manufacturer, as named on the label and in the device documentation, is the only entity that can assign UDIs, receive codes from the issuing entities linked to itself, and register the device in EUDAMED.
Delegation is allowed, liability is not delegable
The paper cites the EU UDI Helpdesk, which confirms that a manufacturer may delegate the practical operation of UDI assignment and application to a third party, for example under a contractual agreement. What cannot move is the responsibility. As the Helpdesk puts it, “the ultimate legal liability for complying with UDI obligations remains with the manufacturer”.
So a service provider or a distributor can carry out the mechanics on the manufacturer’s behalf. The codes still have to be linked to the manufacturer, not to the party doing the work, and the manufacturer still answers for compliance.
What this means for manufacturers
If you sell any device through own-brand-label arrangements, this is worth checking now rather than during an inspection.
Start by mapping every trade name your devices are sold under and identifying who obtained each UDI-DI. Where a distributor holds a code in its own name, that needs correcting: the code should be assigned by you and linked to you at the issuing entity. Then review your EUDAMED registrations under each Basic UDI-DI to confirm every UDI-DI traces back to you as the manufacturer, not to a distributor.
If a third party manages UDI work for you, check that the contractual arrangement reflects the MDCG’s position. Delegating the task is fine. The codes and the registration must still sit with you, and the agreement should make clear that legal responsibility does too.
None of this is a new obligation. The MDCG is clarifying how the existing rules apply, which means enforcement can follow without a transition period. That is the reason to act on it while it is a housekeeping exercise rather than a finding.
If you would like a second pair of eyes on how your UDI-DIs and EUDAMED registrations are structured across brands, get in touch. Untangling an own-brand-label arrangement is far easier before it surfaces in a review than after.