On 29 July 2026 the MHRA published guidance on ambient voice technology-enabled products, the AI scribes now appearing across clinical settings. The guidance sets out when these products qualify as medical devices on the Great Britain market, and how they should be classified if they do.
The short version is that the line is drawn by intended purpose, not by the technology. Two products can run on the same large language model, and one is a regulated medical device while the other is not. What separates them is what the manufacturer claims the product does.
What an AVT product is
Ambient voice technologies are AI tools that capture spoken words and convert them into text or other outputs. In healthcare, they are typically used to support documentation: draft letters, discharge summaries, clinical correspondence and the like. Increasingly, they are built on general-purpose generative AI, which is the source of both their usefulness and the MHRA’s caution.
The regulator’s concern is that flexible technology drifts. A product launched as an administrative note-taker can gain features over time that give it a medical purpose. The guidance is explicit that manufacturers should assess the device regulations with every change to the product, because a modification can move it across the line.
The dividing line: administrative versus medical purpose
The guidance works through nine examples.
Products that stay on the administrative side, and are not medical devices, include an ambient scribe that transcribes a consultation, one that summarises it for clinician review, a tool that formats information into structured data for a clinician to confirm, one that suggests clinical codes based on terms explicitly mentioned in the conversation, and one that drafts a discharge summary or letter for a clinician to edit. The common thread is that the product documents or organises what was said and leaves the clinical thinking to the clinician.
It should be noted that many in the regulatory space have been surprised that the summarisation action is not considered to qualify as a medical device since it is considered that the product would be processing information, giving a result, and the clinician would be following that result thus potentially affecting clinical decision.
Products that cross into being medical devices include a scribe that offers “generated insights” suggesting diagnoses or treatment options, one whose manufacturer claims it “guides diagnosis and treatment planning and improves patient outcomes”, an agent that autonomously orders follow-up tests without clinician review, and a product that analyses data to provide suggested diagnoses for a report. Here the product either does clinical work itself or is marketed as doing so.
Two of those examples are worth dwelling on. In one, the product’s function is purely administrative, but the manufacturer’s marketing claims a medical purpose, and that claim alone makes it a device. In another, a general disclaimer that a feature “is not intended to replace traditional methods of diagnosis” does not save a product whose interface plainly offers diagnostic insights. The MHRA is direct that warnings are the lowest form of risk control, and that manufacturers are expected to design to their stated purpose first.
How these devices get classified
If an AVT product qualifies as a device, classification follows the UK MDR 2002, which uses the classification rules from Annex IX of the old EU Medical Devices Directive. For software, the key rule is Rule 10.
An AVT product that allows direct diagnosis falls into Class IIa. The guidance treats a device as allowing “direct diagnosis” when it provides the diagnosis itself, provides decisive information for making a diagnosis, or is claimed to perform or support a clinician’s diagnostic tasks. Crucially, “decisive information” is read broadly. A product that estimates the probability of a condition from a clinical interaction, where that output is relied on to determine treatment or management, is providing decisive information and should be regulated as Class IIa. Class IIa means conformity assessment by an approved body for UKCA, or a notified body for CE.
A note on scope: GB, and the Northern Ireland difference
This guidance applies to the Great Britain market under the UK MDR 2002. It is not an EU instrument. Under the Windsor Framework, EU MDR continues to apply in Northern Ireland, so a product placed on the NI market is assessed under EU rules and EU classification.
For manufacturers who read this and think in MDR terms, the qualification logic will feel familiar. The MHRA’s reasoning, that intended purpose is set by claims rather than by the underlying model, and that generative AI must be designed to behave within its stated purpose, maps closely onto how the same questions are approached under the MDR. If your software sits anywhere near the boundary between administrative support and clinical decision support, this guidance is a useful test of your own claims regardless of which market you are targeting.
What this means for manufacturers
The fastest route from administrative tool to regulated device is a marketing line that promises clinical benefit. Review your labelling, instructions for use, website and sales materials for any claim that implies diagnosis, treatment, prediction or improved outcomes, and make sure they match what the product is actually intended to do.
Map each feature to an intended purpose. Where a product offers “insights”, recommendations or anything beyond documenting what was said, decide honestly whether that feature carries a medical purpose. If it does, plan for qualification as a device and, most likely under Rule 10, classification as Class IIa with the conformity assessment that entails.
Design against the known hazards of generative AI. The MHRA singles out hallucination as a foreseeable hazard to be mitigated by design rather than covered by a warning. Build that reasoning into your risk management and your technical documentation now, because it is exactly what an assessor will look for.
Deciding whether a feature tips your software into device territory is rarely about the model and almost always about how the product is described and intended. If you are weighing that question, get in touch. It is the kind of qualification and classification work we help clients through regularly, and it is far cheaper to resolve before launch than after.