Executive Insight
The European Commission’s proposal to move the MDR and IVDR from Section A to Section B of Annex I of the EU AI Act appears, at first glance, to be a technical correction but, in practice, it materially reshapes how artificial intelligence used in medical devices and IVDs will be regulated in the EU.
This change means that most AI systems embedded in medical devices and IVDs would no longer be subject to the full horizontal requirements of the AI Act. Instead, regulatory control would be exercised primarily through the existing MDR and IVDR frameworks, with selective and indirect application of AI Act concepts. For manufacturers, this removes a looming layer of duplicative compliance obligations that many had assumed were inevitable.
However, we do not count this as deregulation. The proposal reflects a deliberate policy choice, which is to treat AI in medical devices as a sector-specific risk to be governed within the medical device regime. AI-specific controls will increasingly be embedded via MDR/IVDR GSP requirements, harmonised standards, notified body expectations, and post-market oversight, rather than through the AI Act’s standalone conformity regime.
Manufacturers who interpret this as a simplification without consequence risk misjudging how scrutiny will be applied in practice.
Regulatory Context
Under Regulation (EU) 2024/1689 (the AI Act), Annex I lists Union harmonisation legislation that triggers classification of certain AI systems as high-risk under Article 6(1).
- Section A captures legislation where AI systems are automatically high-risk if they are safety components or regulated products.
- Section B lists legislation where AI relevance is acknowledged, but without automatic application of the AI Act’s Chapter III requirements.
The Commission proposal COM(2025) 1023 explicitly moves Regulations (EU) 2017/745 (MDR) and (EU) 2017/746 (IVDR) from Section A to Section B of Annex I, amending Regulation (EU) 2024/1689 accordingly.
The stated rationale in Recital 23 is to prevent overlaps and simplify the regulatory framework medical devices and IVDs already operate under, with overlaying the full AI Act conformity framework considered structurally duplicative.
Importantly, the proposal preserves regulatory cross-references: when the Commission adopts MDR/IVDR implementing acts, delegated acts, or common specifications affecting devices that qualify as high-risk AI systems, it must take into account Chapter III, Section 2 of the AI Act.
Critical Analysis
What is actually changing
The key change is not whether AI in medical devices is regulated, but where and how that regulation is enforced.
By moving MDR/IVDR to Section B:
- Medical devices and IVDs no longer automatically trigger AI Act Chapter III obligations (risk management systems, data governance, standalone AI quality systems, etc.).
- Manufacturers are no longer required to run parallel conformity assessments under the AI Act for most medical AI applications.
- The notified body, not an AI-specific conformity body, remains the primary gatekeeper.
This resolves a long-standing structural tension: the AI Act was designed for horizontal AI risk, whereas MDR/IVDR regulate clinical risk in a specific context.
What is not changing
The proposal does not eliminate AI-specific scrutiny:
- The Commission explicitly reserves the power to inject AI Act-derived requirements into MDR/IVDR via common specifications or delegated acts.
- Notified bodies designated for devices incorporating high-risk AI systems must still meet AI-related competence requirements under Article 31 of the AI Act.
- Post-market controls, cybersecurity, and vigilance obligations under MDR/IVDR are being strengthened in parallel.
In effect, AI governance is being internalised into the medical device framework, not relaxed.
Practical Implications for Manufacturers
Evidence strategy
AI-related claims will continue to be assessed through:
- Clinical evaluation and performance evaluation,
- Validation of algorithm performance across intended populations,
- Robust justification of training, testing, and update mechanisms.
Manufacturers should not assume that removal from AI Act Chapter III reduces evidentiary expectations.
Timelines
Short-term timelines may improve by avoiding a second conformity pathway. Medium-term timelines may lengthen if there are additional requirements added into the MDR and IVDR to fill any gaps against the AI Act Chapter III requirements.
Strategic Takeaway
The proposed amendment to Annex I of the AI Act is best understood as a “migration” of regulatory authority. For medical devices and IVDs, AI will be governed less by horizontal AI law and more by sector-specific clinical, safety, and lifecycle controls.
This places a premium on disciplined regulatory architecture: integrating AI risk management into existing MDR/IVDR systems rather than treating it as a parallel compliance stream. Manufacturers that align early with this direction will be better positioned to navigate notified body expectations as the regulatory landscape continues to consolidate.
Remember, these are still in the proposal stage, and whether or not these changes are made, expect and prepare that any new requirements introduced in the Act will apply to AI medical devices and IVDs in the future.
Reach out to our team if you have any doubts of how this will impact your existing or upcoming portfolio.