New Harmonised Standards Published in the OJEU: What the 17 June 2026 Updates Mean for Manufacturers

On 17 June 2026, the Official Journal of the European Union published two Commission Implementing Decisions that change the list of harmonised standards behind the Medical Devices Regulation and the In Vitro Diagnostic Medical Devices Regulation. Decision (EU) 2026/1231 amends the MDR list (Implementing Decision (EU) 2021/1182). Decision (EU) 2026/1313 amends the IVDR list (Implementing Decision (EU) 2021/1195).

If you make devices, the practical question is simple. Which of these standards touches your portfolio, and how long do you have to act? Below we walk through every new and amended standard, grouped by family, with the transition dates that matter.

First, a quick reminder of why this list exists. A harmonised standard is voluntary. You do not have to use it. But conformity with a harmonised standard whose reference sits in the Official Journal gives you a presumption of conformity with the matching requirements in the MDR or IVDR. In plain terms, it is the cleanest, best-understood route to showing a notified body that you meet the General Safety and Performance Requirements. When a reference is added, you gain that route. When an old version is scheduled for withdrawal, you need a plan to move across.

The headline change: the EC REP symbol (EN ISO 15223-1)

The single change with the widest reach is the amendment to EN ISO 15223-1:2021, the standard for symbols used with the information a manufacturer supplies. The new amendment, EN ISO 15223-1:2021/A1:2025, does two things. It adds a defined term for “authorised representative”, and it modifies the EC REP symbol so that it is no longer country or region specific. The ‘EC REP’ becomes ‘EU REP’ for the EU:

EU REP

This one matters because it lands on both sides of the regime. It is added under the MDR by Decision 2026/1231 and under the IVDR by Decision 2026/1313. Almost every CE-marked device carries the authorised representative symbol somewhere on its labelling or in its instructions for use, so the practical footprint is enormous.

Both decisions defer the withdrawal of the previous version of EN ISO 15223-1:2021 by a full five years. Under the MDR, the old reference is withdrawn from 15 June 2031. Under the IVDR, it is withdrawn from 17 June 2031. The amended version carries presumption of conformity from the date of publication, 17 June 2026.

What manufacturers should expect. The amended symbol and the new defined term mean a labelling and artwork change across affected products. This is the kind of change that looks small on a single label and becomes a programme of work across a full portfolio, once you account for instructions for use, packaging, translations, controlled documents and existing stock. Five years is generous, but artwork change cycles are slow and tend to bunch up. The sensible move is to fold the new symbol into your next planned artwork revision rather than running a separate relabelling exercise later.

Biological evaluation of medical devices (EN ISO 10993 series)

The MDR decision makes several changes to the biocompatibility family, and this is the most technically significant cluster after the symbols change.

The standout is the newly harmonised EN ISO 10993-1:2025, Part 1, which sets out the requirements and general principles for evaluating biological safety within a risk management process. This is the master standard for biocompatibility. It frames how every other part of the series is applied. Its first appearance on the MDR list gives manufacturers a presumption-of-conformity route aligned to the current edition, and it signals the direction of travel: biological evaluation driven by risk management rather than a fixed test checklist. See our dedicated article on this update to ISO 10993-1.

Three existing parts are updated by amendment, with the previous versions scheduled for withdrawal from 15 December 2027, giving roughly eighteen months of coexistence:

  • EN ISO 10993-23:2021/A1:2025, tests for irritation. The amendment adds further in vitro reconstructed human epidermis models, supporting animal-free irritation testing.
  • EN ISO 10993-12:2021/A1:2025, sample preparation and reference materials.
  • EN ISO 10993-17:2023/A1:2025, toxicological risk assessment of medical device constituents.

One further part is newly added: EN ISO 10993-5:2009 with its amendment EN ISO 10993-5:2009/A11:2025, tests for in vitro cytotoxicity. The A11 amendment is the European annex that ties the standard back to the MDR requirements.

What manufacturers should expect. If you hold biological evaluation plans and reports, expect to refresh them against the 2025 edition of Part 1 and the amended parts. The shift towards reconstructed human epidermis models in Part 23 is good news for anyone moving away from animal testing, but it does mean reviewing your test methods and your justifications. Plan the Part 12, 17 and 23 transitions before 15 December 2027.

Medical electrical equipment (EN 60601 series)

Two changes here. EN 60601-1:2006 with its amendment EN 60601-1:2006/A13:2024, the general requirements for basic safety and essential performance, is newly added to the MDR list. This is the backbone safety standard for electrically powered medical devices, so its presence in the Official Journal gives a large group of manufacturers a clean presumption route against the current consolidated text.

Alongside it, EN IEC 60601-2-83:2020 gains a new amendment, EN IEC 60601-2-83:2020/A1:2025, covering home light therapy equipment. The previous reference is withdrawn from 15 December 2027.

What manufacturers should expect. Electrical device makers should confirm which edition and amendments they currently cite in their technical documentation and align to the harmonised versions. For home light therapy equipment specifically, review the A1:2025 changes and update test evidence where needed ahead of the 15 December 2027 withdrawal of the older reference.

Transfusion equipment (EN ISO 1135 series)

Two newly added standards for single-use transfusion equipment:

  • EN ISO 1135-4:2025, transfusion sets for single use, gravity feed.
  • EN ISO 1135-5:2025, transfusion sets for single use with pressure infusion apparatus.

What manufacturers should expect. Makers of single-use transfusion sets gain a presumption route against the current 2025 editions. If your existing files reference superseded versions, update them to claim the harmonised standards.

Ophthalmic optics (EN ISO 12870 and EN ISO 14889)

Two newly added standards:

  • EN ISO 12870:2025, spectacle frames, requirements and test methods.
  • EN ISO 14889:2025, spectacle lenses, fundamental requirements for uncut finished lenses.

What manufacturers should expect. Spectacle frame and uncut finished lens manufacturers can now point to harmonised 2025 editions. This is a straightforward alignment exercise for most, but worth confirming your test reports map to the current requirements.

Non-active surgical implants: mammary implants (EN ISO 14607)

EN ISO 14607:2025, mammary implants, specific requirements, is newly added to the MDR list. Breast implants are Class III devices, so the harmonised standard carries particular weight for demonstrating conformity at the highest risk class.

What manufacturers should expect. Mammary implant manufacturers should align their technical documentation to the 2025 edition. Given the device class and the level of notified body scrutiny involved, treat this as a priority alignment rather than a routine update.

Washer-disinfectors (EN ISO 15883 series)

Four newly added standards covering reprocessing equipment:

  • EN ISO 15883-1:2025, general requirements, terms and definitions and tests.
  • EN ISO 15883-2:2025, thermal disinfection for critical and semi-critical medical devices.
  • EN ISO 15883-3:2025, thermal disinfection for human waste containers.
  • EN ISO 15883-7:2025, chemical disinfection for non-critical thermolabile medical devices and healthcare equipment.

What manufacturers should expect. Manufacturers of washer-disinfectors gain harmonised references across the core parts of the series. If you supply reprocessing equipment, review which parts apply to your product and update your conformity claims to the 2025 editions.

Prosthetics (EN ISO 22675)

EN ISO 22675:2025, testing of ankle-foot devices and foot units, requirements and test methods, is newly added.

What manufacturers should expect. Manufacturers of ankle-foot prosthetic devices and foot units can claim the harmonised 2025 edition. Confirm your mechanical test evidence aligns with the current method.

Sharps injury protection (EN ISO 23908)

EN ISO 23908:2025, sharps protection mechanisms for single-use needles, introducers for catheters and needles used for blood testing, monitoring, sampling and medical substance administration, is newly added.

What manufacturers should expect. Makers of safety-engineered sharps and protection mechanisms gain a harmonised route. Align your design and test documentation to the 2025 edition.

The IVDR side: one change, broad reach

The IVDR decision, 2026/1313, makes a single change, and it is the symbols standard discussed above. EN ISO 15223-1:2021/A1:2025 is added under the IVDR, with the previous version withdrawn from 17 June 2031. For IVD manufacturers, the takeaway is the same as on the MDR side: plan the EC REP symbol and authorised representative term into your next labelling revision well ahead of the 2031 deadline.

What to do next

Three steps will cover most manufacturers. Map these standards against your product portfolio and your current technical documentation. Identify which changes are amendments with a withdrawal date, the 15 December 2027 and 2031 deadlines being the ones to diary, and which are newly added references that simply give you a cleaner conformity route. Then schedule the work, starting with anything that drives labelling or artwork change, since those take the longest to ripple through a portfolio.

If you would like a second pair of eyes on which of these standards apply to your devices and how to phase the transitions, that is exactly the kind of work we do. Let’s talk it through.

Sources

Commission Implementing Decision (EU) 2026/1231

Commission Implementing Decision (EU) 2026/1313