Placing SaMD on the EU Market: Where Software Developers Start

You have built software that does something genuinely useful in a clinical setting. It flags a result, supports a diagnosis, helps plan a treatment, or monitors a patient. At some point the question lands: how do we put this on the European market without getting it wrong?

This is a foundation piece for software developers and product teams who are new to the EU’s medical device rules. We are going to walk through the product side of compliance, the requirements that attach to the device itself and the evidence you build around it. We are deliberately leaving out the broader company obligations such as the EU authorised representative, importer and distributor duties, and general quality system setup, except where the quality system touches design and development directly. Those topics deserve their own posts.

We are also keeping this product agnostic. Artificial intelligence, machine learning and the EU AI Act bring their own layer of requirements, and we will cover those separately. The principles below apply whether or not your software learns.

First, the language: SaMD, MDSW and what counts as a device

Two terms get used almost interchangeably. Software as a Medical Device (SaMD) is the international term from IMDRF, describing software intended for a medical purpose that performs that purpose without being part of a hardware medical device. Medical Device Software (MDSW) is the term the EU legislation and guidance uses, and it is slightly broader: it covers software that drives or influences a hardware device as well as standalone software. For most of this post the distinction does not change what you have to do.

Before any of the product requirements apply, your software has to actually be a medical device. That step is called qualification, and it turns on intended purpose. If the software is intended by you, the manufacturer, for a medical purpose such as diagnosis, prevention, monitoring, prediction, prognosis or treatment, it is very likely in scope. The reference document here is MDCG 2019-11, the guidance on qualification and classification of software under the MDR and IVDR, updated to Revision 1 in June 2025. It is the right place to start, and it includes decision diagrams that walk you through qualification before you ever reach classification.

One honest word of caution. Qualification is where a lot of products are misjudged, usually in the optimistic direction. “It is just a wellness app” or “it only displays data” are claims that need to survive a deep look at MDCG 2019-11 and your own intended purpose statement. Get this wrong and everything downstream is built on sand.

Classification: Rule 11 and why almost everything lands in Class IIa

Once your software qualifies as a medical device, the next step is to classify it. Software is an active device, and under the EU MDR the relevant classification rule is Rule 11 in Annex VIII. It reads:

Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa, except if such decisions have an impact that may cause:

– death or an irreversible deterioration of a person’s state of health, in which case it is in class III; or

– a serious deterioration of a person’s state of health or a surgical intervention, in which case it is classified as class IIb.

Software intended to monitor physiological processes is classified as class IIa, except if it is intended for monitoring of vital physiological parameters, where the nature of variations of those parameters is such that it could result in immediate danger to the patient, in which case it is classified as class IIb.

All other software is classified as class I.

Rule 11 is arguably the most deceptive. The phrases “information used to take decisions” and “serious deterioration” carry enormous weight, and the rule is the one most teams wrestle with. In practice, many devices that feel low risk to their developers land in Class IIa, because almost any software that informs a clinical decision meets that first sentence. A genuine Class I medical device software product is rare, and if you have concluded you have one, it is worth a second look.

You do not have to interpret Rule 11 alone. MDCG 2021-24, the guidance on classification of devices under the MDR, works through Annex VIII rule by rule and gives worked examples for each class. Read it alongside MDCG 2019-11 and the regulation itself. The classification you land on drives the conformity assessment route and how much notified body involvement you need, so it is worth the care. Build and record your classification rationale and ensure it is watertight before moving on.

A note for IVD software

If your software is an in vitro diagnostic, you follow the IVDR rather than the MDR, and classification works differently. There is no Rule 11 or software-specific rule. IVDs are classified into Classes A to D using the rule set in IVDR Annex VIII, and software is generally classified according to the information or test it drives. MDCG 2019-11 covers both regulations, so it remains your starting point. The rest of this post applies to IVD software too, with performance evaluation standing in for clinical evaluation, which we indicate below.

A proposal to rewrite Rule 11

Rule 11 has been criticised since the MDR came into force for pushing almost every piece of decision-support software to Class IIa or higher, leaving genuine Class I software all but extinct. The European Commission’s 2026 proposal to amend the MDR takes this on directly by rewriting Rule 11 from scratch. It is a proposal, not yet law, so today’s classification still follows the wording set out above. It is worth understanding now, because if adopted it changes the starting point for almost every software device.

The proposed wording reads:

Software which is intended to generate an output that confers a clinical benefit and is used for diagnosis, treatment, prevention, monitoring, prediction, prognosis, compensation or alleviation of a disease or condition is classified as class I, unless the output is intended for a disease or condition:

  • in a critical situation with a risk of causing death or an irreversible deterioration of a person’s state of health, in which case it is classified as class III;
  • in a serious situation with a risk of causing a serious deterioration of a person’s state of health or a surgical intervention, or to drive clinical management in a critical situation in which cases it is classified as class IIb;
  • in a non-serious situation, or to drive clinical management in a serious situation or to inform clinical management in a critical or serious situation in which cases it is classified as class IIa.

Two things change in substance. First, the default flips. Software starts at Class I and only climbs when the clinical stakes justify it, the reverse of the current presumption that most software sits at Class IIa or above. Second, the rule borrows the structure international regulators already use through the IMDRF framework, classifying on two axes at once: the seriousness of the healthcare situation or condition, described as critical, serious or non-serious, and the significance of what the software does with the information, namely whether it generates the clinically beneficial output itself, drives clinical management, or informs it.

If this proposal is accepted, the practical effect will be quite extensive. Some products that are Class IIa under today’s Rule 11 could sit at Class I, which changes whether a notified body has to be involved at all. The flip side is that the new language leans heavily on terms such as “drive” and “inform” clinical management, and “critical”, “serious” and “non-serious” situations, all of which will need careful, evidenced interpretation against your intended purpose. We would expect updated guidance to follow, much as MDCG 2021-24 supports the current rule. Until the amendment is adopted, classify against the rule in force, and factor the proposed direction into any roadmap that stretches into the next couple of years.

The General Safety and Performance Requirements: the real backbone

It is tempting to treat compliance as a list of standards to tick off. The more accurate picture is that the regulation sets out the General Safety and Performance Requirements (GSPRs) in Annex I of the MDR and IVDR, and the standards are simply the recognised ways of meeting them. Your technical documentation has to show, requirement by requirement, that the device conforms.

For medical device software, the requirements that matter most sit in MDR Annex I, Chapter II, Section 17, covering electronic programmable systems and software. In summary, these require that software be developed and manufactured in accordance with the state of the art, taking into account the principles of the development life cycle, risk management, information security and verification and validation. Section 17 also requires you to set out the minimum requirements for the hardware, IT network characteristics and IT security measures the software needs to run as intended. Beyond Section 17, pay attention to the general requirements in Sections 1 to 9, the labelling and instructions for use requirements in Section 23 (electronic instructions for use are often relevant for software), and repeatability, reliability and performance expectations throughout.

The practical move is to build a GSPR checklist, list each applicable requirement, and map it to the evidence and the standard you are using to satisfy it. That checklist becomes the backbone of your compliance.

The software life cycle: IEC 62304

This is the standard most developers meet first, and the one notified bodies expect you to follow thoroughly. IEC 62304 defines the life cycle processes for medical device software, covering development, maintenance, risk management, configuration management and problem resolution. It applies to both standalone software and to software embedded in a hardware device.

IEC 62304 introduces its own software safety classification, Classes A, B and C, in order of increasing risk, based on the severity of harm that could result if the software failed. The classification then determines which activities in the standard you must perform. So the first step is to classify the software using the standard’s process, then work through the remaining clauses to see what applies to you. A useful and often overlooked point: this classification can be applied at the level of software items or components, not just the whole product, which can reduce the burden where lower risk components are clearly separated.

IEC 62304 does not sit on its own. The software development plan sits within, or alongside, the overall device development plan, which in turn has to meet the design and development requirements of ISO 13485, the quality management system standard. In other words, IEC 62304 supplements the ISO 13485 design and development process rather than replacing it. Your software requirements specification feeds the design inputs, and outputs such as the software architecture, the list of SOUP (software of unknown provenance, typically third party and off the shelf components), and user documentation flow out the other side.

A reality we see often: clients arrive at the end of development, with a working product and no life cycle documentation. Compliance is still achievable. It means reconstructing the development history into the structure IEC 62304 expects and demonstrating to the assessor that the process meets the standard. It is more work done in reverse than it would have been done in order, which is the honest argument for engaging with the standard early.

Standalone software products: IEC 82304-1

If your product is software only, meaning standalone software intended to run on general computing equipment that was not built specifically for it, such as a mobile app, a web application or a cloud based tool, then IEC 82304-1 applies as well. It is frequently overlooked, and that is a costly oversight. Its full title is “Health software, Part 1: General requirements for product safety”, and it addresses both the safety and the security of software only products across their whole life.

The cleanest way to picture the relationship between the two software standards is a division of labour. IEC 62304 governs how you engineer the software, the development and maintenance processes inside your organisation. IEC 82304-1 sits on top and deals with the finished product you actually place on the market. The standard itself states that it relies heavily on IEC 62304 for the software development process, so the two interlock rather than compete. You use IEC 62304 for the engineering, and IEC 82304-1 to wrap the product, its requirements, its validation and its accompanying information into something you can defend as safe in real use.

In practice, IEC 82304-1 walks you through a product life cycle that covers:

  • Product use requirements, capturing what the intended users need and the environment the software will really run in, and then verifying those requirements are complete and correct.
  • System requirements, including the platform, hardware and IT environment the product depends on to operate safely.
  • Product validation, confirming the finished product meets its use requirements in the intended use environment, rather than only that the code passed its tests.
  • Identification and accompanying documents, the information you ship with the product so it can be installed, configured and used safely and securely.
  • Post-market activities, including re-validation when the product changes, ongoing communication with users, and decommissioning and disposal at end of life.

This is where standalone SaMD teams most often have a gap. You can follow IEC 62304 to the letter and still have said nothing about the messy reality your software runs in: someone else’s phone, an operating system you do not control, a hospital network, a browser that updates without asking. IEC 82304-1 is where you make those platform assumptions and the minimum system requirements explicit. That work feeds straight into the General Safety and Performance Requirements in MDR Annex I, which ask you to state the hardware, IT network and IT security conditions your software needs to perform as intended. The version adopted in Europe is EN 82304-1:2017, which corresponds to IEC 82304-1:2016.

Risk management: ISO 14971

Risk management is not a software specific activity, but it sits underneath everything above. ISO 14971 is the standard for applying risk management to medical devices, and it is the engine that feeds both your GSPR conformity and your IEC 62304 software safety classification. Your hazard analysis identifies what can go wrong, the severity of the resulting harm and the controls you put in place, and that analysis directly informs how rigorously you have to develop the software. Build the risk management file early; almost every other document references it.

Usability and human factors: IEC 62366-1

Software is used by people, often under pressure, sometimes by patients themselves. Use error is a recognised source of harm, so usability engineering is part of the picture. IEC 62366-1 sets out the usability engineering process for medical devices, linking how the user interacts with your software to the risk management file. For SaMD with clinical decision support, monitoring dashboards or patient facing interfaces, this is not optional polish; it is evidence that the design accounts for how it will really be used.

Cybersecurity: now a core product requirement

For medical device software, security and safety are tied together. A vulnerability that lets someone alter a result, block access or corrupt data is a safety problem, and the regulation treats it as one. MDR Annex I, Section 17, makes this explicit. Software must be developed according to the state of the art taking information security into account (17.2), and you must define the minimum requirements for the IT environment and the IT security measures that protect against unauthorised access (17.4), alongside the repeatability, reliability and performance expectations in 17.1. The IVDR carries equivalent requirements in its own Annex I, Section 16. Security is therefore part of your GSPRs and has to be evidenced in the technical documentation from the start.

The central guidance is MDCG 2019-16 on cybersecurity for medical devices. It sets out a security by design and security by default approach, and it treats security as a whole life cycle concern that splits into two halves. Before market, you run a security risk assessment that sits alongside your ISO 14971 safety risk management, threat model the product, and design in controls such as authentication, encryption and a secure update mechanism. After market, you monitor for new vulnerabilities, patch them, communicate with users, and feed serious issues into incident reporting. The safety risk file and the security risk assessment have to talk to each other constantly, because a new security weakness usually surfaces as a new safety hazard.

The emerging state of the art standard is IEC 81001-5-1 (adopted in Europe as EN IEC 81001-5-1:2022), covering security activities across the health software life cycle. It is process oriented rather than prescriptive, so it tells you which activities to build into development and maintenance rather than which encryption library to choose, and it dovetails with IEC 62304. Treat it as the benchmark assessors will increasingly expect. Alongside it, a few artefacts are becoming standard expectations: a software bill of materials (SBOM) listing your third party and SOUP components so you can track their vulnerabilities, a documented vulnerability handling and coordinated disclosure process, and a secure, verifiable mechanism for shipping updates.

One forward look is worth noting. Medical devices placed on the market under the MDR and IVDR remain outside the product requirements of the EU Cyber Resilience Act (Regulation (EU) 2024/2847), which avoids regulating the same product twice, and the Commission’s 2026 proposal to amend the MDR keeps that exclusion in place. What the proposal does add is a new Article 87a that aligns cybersecurity reporting between the two regimes. Under it, manufacturers would report any actively exploited vulnerability, as defined in Article 3(42) of the Cyber Resilience Act, and any severe incident under Article 14(5) of that Act affecting the security of the device, to the national computer security incident response teams (CSIRTs) and to ENISA, within 30 days, through the MDR’s own electronic system. The direction of travel is alignment of incident reporting, rather than a wholesale move of devices into the Cyber Resilience Act. It is also worth remembering that any non-medical companion products you sell, such as a wellness app or a fitness tracker, can fall under the Cyber Resilience Act in their own right.

Clinical evaluation, or performance evaluation for IVDs

Demonstrating that your software is safe and that it works as intended requires clinical evidence, even for software. The guidance is MDCG 2020-1, covering clinical evaluation under the MDR and performance evaluation under the IVDR for medical device software. It frames the evidence around three building blocks: the valid clinical association between your software’s output and the targeted clinical condition, the technical performance showing the software correctly processes inputs into outputs, and the clinical performance showing the output achieves the intended purpose in the target population. For IVD software, the equivalent is the performance evaluation built on scientific validity, analytical performance and clinical performance. Planning this evidence early shapes what you need to collect during development, so it is not a step to leave until the end.

Pulling it together: the technical documentation

All of the above lands in one place, your technical documentation, set out in Annex II of the MDR (Annex II of the IVDR), with post market surveillance (PMS) documentation in Annex III. This is the file a notified body assesses for higher risk classes, and that you hold and keep current for all classes. For software, a coherent file ties together the device description and intended purpose, the GSPR checklist with evidence, the risk management file, the IEC 62304 life cycle documentation, the IEC 82304-1 product requirements where relevant, usability and cybersecurity evidence, verification and validation reports, and the clinical or performance evaluation.

A few software specific items are easy to miss:

  • UDI for software. MDCG 2018-5 explains how unique device identification applies to software, including when a new UDI-DI is triggered by significant changes such as modified algorithms, architecture or user interfaces.
  • Software combined with hardware. If your app relies on a sensor, wearable or other hardware component, MDCG 2023-4 covers how those combinations are treated and placed on the market.
  • Apps on online platforms. If you distribute through app stores, MDCG 2025-4 sets out how to make MDSW apps available safely on online platforms, including the information that needs to reach users.

Common mistakes we see

  • Underclassifying at qualification. Deciding the product is not a medical device, or is only Class I, without working through MDCG 2019-11 and Rule 11 with MDCG 2021-24.
  • Treating standards as the goal. Following IEC 62304 to the letter while never mapping back to the GSPRs in Annex I, which is what conformity actually rests on.
  • Leaving documentation to the end. Reconstructing a life cycle history retrospectively is achievable, but it costs far more than building it as you go.
  • Forgetting IEC 82304-1. Standalone software teams often stop at IEC 62304 and miss the product level requirements.
  • Bolting on cybersecurity and usability late. Both feed the risk management file and both are expected from the start.

A sensible order to work in

If you are starting from scratch, a workable sequence is: qualify the device against intended purpose, classify it under Rule 11 with the guidance, open your risk management file, stand up the IEC 62304 life cycle within your ISO 13485 design and development process, layer in IEC 82304-1 if you are software only, address usability and cybersecurity as you design, plan the clinical or performance evidence in parallel, and assemble it all into the Annex II technical documentation against a GSPR checklist. None of these are truly sequential; they inform each other, which is exactly why starting early matters.

A note on AI-enabled devices

If your software uses artificial intelligence or machine learning, everything above still applies. AI does not exempt you from qualification, classification, the GSPRs, the life cycle standards, risk management or clinical evaluation. It adds a layer on top, and that layer is involved enough to deserve its own post, which we will publish separately.

As things stand, an AI medical device can be dual-regulated. It is a medical device under the MDR or IVDR, and where it requires a notified body for conformity assessment it is also a high-risk AI system under the EU AI Act (Regulation (EU) 2024/1689), through Article 6(1) and Annex I. That brings a further set of obligations covering data and data governance, transparency, human oversight, accuracy, robustness, logging and risk management for the AI itself. Following the 2026 Digital Omnibus agreement, the high-risk obligations for products in this category are now expected to apply from 2 August 2028, which buys time but does not remove the work.

There is also movement on whether AI medical devices remain within the AI Act’s high-risk scope at all. The 2026 simplification proposals would carve medical devices out of the AI Act’s high-risk requirements and have them assessed under the MDR and IVDR instead, on the logic that running two parallel conformity regimes for one product is wasteful. That sounds like a reprieve, and in administrative terms it may be. Our expectation, though, is that the substance does not disappear. The data governance, transparency, human oversight and robustness expectations are most likely to return through harmonised standards and guidance under the MDR, which is how the framework usually absorbs requirements of this kind. So if you are building AI now, build to those principles regardless of which regulation formally owns them. We will go through all of this properly in the dedicated AI post.

Where Specculo comes in

This is the foundation. Each layer, the life cycle under IEC 62304, the standalone product requirements under IEC 82304-1, clinical evaluation, cybersecurity, and the AI specific requirements that sit on top for learning systems, go deeper than we can achieve in a single post.

If you are mapping your route to the EU market and want a second pair of eyes on classification, your technical file, or the gap between where your documentation is and where a notified body needs it to be, that is the work we do alongside software teams every day. Let’s talk it through. No pressure, just a clear view of your path.