Introduction
Since 2001, Switzerland has regulated medical devices in alignment with the EU, integrating into the European internal market and market surveillance system through a Mutual Recognition Agreement (MRA) with the EU. This MRA facilitated close regulatory cooperation, allowing mutual recognition of conformity assessments so that devices approved in the EU could be marketed in Switzerland without additional evaluation, and vice versa.
However, as the EU Medical Device Regulation (MDR) implementation deadline approached on 26 May 2021, negotiations to update the MRA failed. Consequently, the EU declared the previous MRA invalid, designating Switzerland as a third country. This change led to significant impacts:
- Loss of Swiss access to the EU’s Eudamed database,
- Suspension of cooperation on EU market surveillance,
- Swiss authorities now relying exclusively on domestic processes for device registration, vigilance, and post-market surveillance.
While Swiss regulations remain largely harmonised with the EU’s MDR and IVDR, they now function independently. The updated Swiss Medical Devices Ordinance (MedDO), effective 26 May 2021, and the updated Ordinance on In Vitro Diagnostic Medical Devices (IvDO), effective 26 May 2022, implement much of the EU MDR/IVDR framework but have a few uniquely Swiss twists that will be explained below.
Swiss Authorised Representative (CH REP)
Manufacturers not established in Switzerland must appoint a Switzerland-based Authorised Representative (CH REP) before placing devices on the Swiss market. This requirement applies irrespective of whether the manufacturer is located within the EU/EEA.
So, who is a CH REP?
A CH REP is a natural person or legal entity domiciled in Switzerland, formally mandated by the manufacturer to act as the official regulatory liaison with Swiss authorities. The CH REP assumes key legal responsibilities defined under MedDO and IvDO, ensuring compliance and communication on behalf of the manufacturer.
Key Differences Between MedDO/IvDO and EU MDR/IVDR
While Switzerland’s regulations align closely with EU law, there are a few important differences worth noting:
- Local Representation: A Swiss-based Authorised Representative is mandatory for non-Swiss manufacturers, as already described.
- Registration Numbers: A Swiss Single Registration Number (CHRN) is issued to economic operators (manufacturers, authorised representatives, importers), differing from the EU’s Single Registration Number (SRN).
- Importer Requirement: Foreign devices placed on the Swiss market require a designated importer based in Switzerland.
- Databases: Instead of Eudamed, Switzerland runs its own database called Swissdamed.
- Language: Device labels and instructions must be provided in German, French, and Italian, except under specific exemptions (e.g., professional use only), when placed on the Swiss Market.
- Vigilance Reporting: Incidents are reported directly to Swissmedic, using Swissmedic forms, not via Eudamed.
- Hospital Vigilance: Swiss hospitals must maintain a Quality Management System, designate a vigilance contact person, and keep records for 15 years; requirements not mirrored in EU law.
- PSUR & SSCP: Manufacturers must submit PSURs directly to the designated body and publish SSCP themselves, as Swissdamed does not handle publication.
Swissdamed: The Swiss Medical Device Database
Swissdamed is the new medical devices information system operated by Swissmedic (the Swiss authority responsible for the authorisation and supervision of therapeutic products). Due to lost access to Eudamed, Swissdamed is being introduced in phases:
- Actors Module (launched August 2024): Registers economic operators and issues CHRN.
- Devices Module (expected 2025–2026): Will enable device registrations, initially voluntary some time in 2025 and becoming mandatory, starting 1 July 2026.
Device Registration Timelines:
- Products, systems, and procedure packs placed on the market after 1 July 2026 must be registered in Swissdamed before being placed on the market, with a transitional period lasting until 31 December 2026.
- Those placed on the market before 1 July 2026, but still available after 31 December 2026, must be registered by 31 December 2026.
- Similarly, products, systems, and procedure packs placed on the market at the start of the registration requirement (1 July 2026) but no longer available after 31 December 2026, must also be registered by 31 December 2026.
Note: The device registration requirement applies for devices under the current legislation (MedDO, IvDO) as well as devices complying with the old legislation which are still being made available on the market.
Actor Registration Process on Swissdamed
1. Company Registration
- The process begins by logging in to Swissdamed using a CH-LOGIN account.
- The company must then be registered via Zefix (the Swiss commercial registry). This registration is done directly through Swissdamed.
- A verification code will be sent by post to the company’s address, usually within a week. Once that code is entered on Swissdamed, the company is officially registered.
- The user who does the registration automatically becomes the Company Admin and can invite other users or manage permissions.
2. Actor Registration
After the company is set up, actor registration can take place:
- The registering user will automatically be assigned the role of Actor Admin, with the ability to manage actor-related data and users.
- While specific required details aren’t fully outlined, the process appears to involve:
- Economic operator details.
- Contact and invoice information.
- PRRC details.
- Upload of an excerpt from the commercial register (not older than three months).
- If it’s a natural person registering, a valid residential address registration document (not older than three months) is also required.
3. Application Review and Approval
- After submission, Swissmedic typically processes the application within 30 days.
- A notification via email will be sent once the application is approved or rejected.
4. Fees
- There’s a flat CHF 200 fee for issuing the CHRN.
- Invoices are issued only to addresses in Switzerland or Liechtenstein and are sent once the CHRN is provided.
- Fees also apply even if the application gets rejected.
Device Registration on Swissdamed
The registration of devices is currently still in development and is not yet available on Swissdamed. However, certain medical devices must be notified to Swissmedic via a separate notification process. You can check out the details here: Notification of medical devices.
Switzerland–Liechtenstein Customs Union
Due to the customs treaty between Switzerland and Liechtenstein, the terms “established in Switzerland” and “on the market in Switzerland” extend to include Liechtenstein. Devices placed on the market in accordance with MedDO/IvDO cover both countries. Notably, manufacturers based in Liechtenstein are not required to appoint a Swiss Authorised Representative.
How Specculo Can Help
The shift in Switzerland’s status to a “third country” under EU MDR has created new obligations for non-Swiss manufacturers: from appointing a Swiss Authorised Representative (CH REP) to navigating the Swissdamed database and unique MedDO/IvDO requirements. Specculo can help you enter and remain compliant in the Swiss market with services tailored to these changes.
We can:
- Engage a CH REP – Serving as your official liaison with Swissmedic, ensuring ongoing compliance, handling communications, and maintaining your documentation.
- Guide Swissdamed registrations – Managing your company and actor registration process, obtaining your CHRN, and preparing for device module requirements ahead of the 2026 deadlines.
- Coordinate importer arrangements – Ensuring you meet Swiss-based importer requirements and helping establish compliant supply chain partners.
- Review labelling and IFU – Verifying that your product information meets Switzerland’s multilingual requirements (German, French, Italian) and any exemptions.
- Support vigilance and PMS – Assisting with Swiss-specific vigilance reporting to Swissmedic and aligning your post-market activities with MedDO/IvDO expectations.
- Regulatory strategy and training – Providing tailored advice and updates to keep your team informed of evolving Swiss requirements.
By partnering with Specculo, you gain an experienced regulatory team that understands both the EU MDR and Switzerland’s unique framework, allowing you to streamline compliance, avoid delays, and maintain uninterrupted market access in Switzerland and Liechtenstein.